User Manual
87 
CMC? 
Description:     
Queries the measured/readback current output 
value of the selected power supply 
Query: 
CMC? 
Return 
String: 
1.234 A 
The selected power supply has a measured output 
current of 1.234 A. 
CDVC? 
Description:     
Queries the display voltage and current of the 
selected power supply. 
Example: 
CDVC? 
Queries channel 1   
Return 
String: 
100.23,1.567 
Display voltage is 100.23 V and current is 1.567 A. 
Format: 
<voltage>,<current> 
COUT 
Description:     
Configures/Queries the output state of the selected 
power supply. 
Example: 
COUT ON 
Sets the selected power supply’s output to ON. 
COUT? 
Queries the selected power supply’s output state. 
COV 
Description:     
Configures/Queries the overvoltage protection 










