Datasheet

Battery Analyzers
BA6010 Series
4
www.bkprecision.com
Specifications
Voltage Measurement (BA6010)
SLOW, MED
Range Maximum Display Value Resolution Accuracy
6 V 6.5000 V 100 µV
±(0.05 % FS)
60 V 65.000 V 1 mV
FAST
Range Maximum Display Value Resolution Accuracy
6 V 6.500 V 1 mV
±(0.1 % FS)
60 V 65.00 V 10 mV
Voltage Measurement (BA6011)
SLOW, MED
Range Maximum Display Value Resolution Accuracy
30 V 35.000 V 1 mV
±(0.05 % FS)
300V 310.00 V 10 mV
FAST
Range Maximum Display Value Resolution Accuracy
30 V 35.00 V 10 mV
±(0.1 % FS)
300 V 310.0 V 100 mV
Model BA6010, BA6011
Measurement
Parameters
Main V, R
Auxiliary
L, C, D, Z, X, Q, θd, and θr
Test Frequency 1 kHz ± 0.2 Hz
Display Resolution 5 digits (SLOW & MED), 4 digits (FAST)
Measurement Speed
SLOW, approx. 6.25 measurements/sec
MED, approx. 10 measurements/sec
FAST, approx. 50 measurements/sec
Temperature
Coefficient
Voltage Meas. 0.005 % / ºC
Resistance Meas. 0.05 % / ºC
Triggering Internal, External, Manual, Bus
Delay Time On / Off, 0 ms to 60 s
Averaging 1 to 255 samples
Statistical Calculations
Valid data count, Invalid data count, Mean, Maximum, Minimum,
Standard Deviation, Sample Standard Deviation,
Process Capability Index (Dispersion), Process Capability Index (Deviation)
All specifications apply to the unit after a temperature stabilization time of 15 minutes over an ambient temperature
range of 20 °C ± 5 °C. Specifications are subject to change without notice.