Datasheet

Battery Analyzers
BA6010 Series
3
www.bkprecision.com
Flexible operation
On screen monitoring system detects test probe
contact failure and damaged leads for reliable
measurements.
Error: HI drive open
Error: LO drive open
Error: HI sense open
Error: LO sense open
Error: Measure line open
Comparator function
The comparator function evaluates measurements
against a user specified upper and lower limit for
pass/fail (Go/No Go) style testing. Comparative
evaluations can be made using primary, secondary
or both measurements. The front panel PASS /
FAIL indicator will illuminate and a sounder can
be enabled for audible confirmation.
Compare test - below limits
Compare test - above limits
Bin comparator display
Statistical function
Statistical tools menu
The analyzers can perform statistical calculations
on the measurements and display the results
on-screen.
Kelvin clip test fixture
The 4-terminals on the front of the BA6010
Series are used together with the Kelvin clip test
fixture. This system minimizes the influence of
the test lead resistance and improves
measurement accuracy.
Improved measurement accuracy
4-wire kelvin connection
Bins for sorting devices
Binning function
Quickly sort components using up to 9 bins. The
bin results are displayed on-screen with each
cycle. The handler interface includes dedicated
signal pins for each bin, Pass/Fail and end of
measurement. The handler interface is suitable
for integration with device handler systems or
programmable logic controllers (PLC) used in
production automation.
Trace function
Trace display
The trace function samples and plots two
user-selected measurement readings over a
specified time. Enable cursors for viewing plotted
values and time stamp information.
Cursor with
measurements