User manual
94
Figure 5.1 – Voltage Calibration Diagram
2. Set the DMM to a DC voltage measurement. Select the
VOLTAGE parameter in the CAL menu and press . The
following will be displayed:
The P: 1/4 signifies there are 4 points to be calibrated, and the
current calibration point is 1.
3. Input the voltage reading from the DMM in the METER
parameter by using the numeric keypad and press .
4. Repeat this procedure by entering the meter readings for points
2 – 4.
5.2 Current Calibration