Model: PVS10005, PVS60085, PVS60085MR High Power Programmable DC Power Supply USER MANUAL
Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument. Before applying power to this instrument: Read and understand the safety and operational information in this manual. Apply all the listed safety precautions. Verify that the voltage selector at the line power cord input is set to the correct line voltage.
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected to the mains supply. The voltages in the environment are typically derived from a limited-energy transformer or a battery. Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the mains supply at a standard wall outlet or similar sources. Example measurement environments are portable tools and household appliances.
ensure it is appropriate for your country of use. Ground the Instrument To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This instrument is grounded through the ground conductor of the supplied, three-conductor AC line power cable. The power cable must be plugged into an approved threeconductor electrical outlet. The power jack and mating plug of the power cable meet IEC safety standards. Do not alter or defeat the ground connection.
In the presence of noxious, corrosive, or flammable fumes, gases, vapors, chemicals, or finely-divided particulates. In relative humidity conditions outside the instrument's specifications. In environments where there is a danger of any liquid being spilled on the instrument or where any liquid can condense on the instrument. In air temperatures exceeding the specified operating temperatures.
instrument from service, label it as not to be operated, and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the instrument. Clean the instrument only as instructed Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual.
after making the measurements by testing with known-operating voltage sources and test for both DC and AC voltages. Do not attempt any service or adjustment unless another person capable of rendering first aid and resuscitation is present. Do not insert any object into an instrument's ventilation openings or other openings. Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit exists.
This instrument contains one or more cooling fans. For continued safe operation of the instrument, the air inlet and exhaust openings for these fans must not be blocked nor must accumulated dust or other debris be allowed to reduce air flow. Maintain at least 25 mm clearance around the sides of the instrument that contain air inlet and exhaust ports. If mounted in a rack, position power devices in the rack above the instrument to minimize instrument heating while rack mounted.
Compliance Statements Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with separate collection systems) This product is subject to Directive 2002/96/EC of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste.
CE Declaration of Conformity This instrument meets the requirements of 2006/95/EC Low Voltage Directive and 2004/108/EC Electromagnetic Compatibility Directive with the following standards.
Safety Symbols Refer to the user manual for warning information to avoid hazard or personal injury and prevent damage to instrument. Electric Shock hazard On (Supply). This is the AC mains connect/disconnect switch on the front of the instrument. Off (Supply). This is the AC mains connect/disconnect switch on the front of the instrument.
CAUTION indicates a hazardous situation which, if not avoided, will result in minor or moderate injury WARNING indicates a hazardous situation which, if not avoided, could result in death or serious injury DANGER indicates a hazardous situation which, if not avoided, will result in death or serious injury.
Table of Contents Safety Summary ................................................................... i Compliance Statements ...................................................................... viii Safety Symbols ....................................................................................... x 1 General Information ..................................................... 1 1.1 Product Overview....................................................................... 1 1.2 Package Contents ........
2.4 Preliminary Check .................................................................... 15 Warm-up Time ................................................................................ 15 Output Check .................................................................................. 16 Check Model and Firmware Version ............................................... 17 3 Front Panel Operation ................................................ 19 3.1 Menu Options ........................................
3.7 PV Simulation ........................................................................... 57 3.8 Program Function ..................................................................... 60 Overview ......................................................................................... 60 Configure Program Parameters ...................................................... 61 3.9 Timer Function ......................................................................... 65 3.10 Slew Rate Setting .......
6 5.5 External Voltage Programming Calibration .............................. 98 5.6 External Current Programming Calibration.............................. 99 5.7 CC Calibration of External Voltage ......................................... 100 5.8 CC Calibration of External Current ......................................... 101 Troubleshooting Guide ............................................. 103 General .........................................................................................
1 General Information 1.1 Product Overview B&K Precision models PVS60085, PVS60085MR, and PVS10005 are high voltage programmable DC power supplies with single outputs that offer the maximum power output up to 5100 watts (0 - 600 V / 8.5 A or 0 1000 V / 5 A). By connecting up to 4 power supplies in parallel, a maximum output power can reach up to 20.4 kW. These power supplies are fully programmable and controllable through analog programming, USB, RS232, RS485, GPIB and Ethernet interface.
1.2 Package Contents Please inspect the instrument mechanically and electrically upon receiving it. Unpack all items from the shipping carton, and check for any obvious signs of physical damage that may have occurred during transportation. Report any damage to the shipping agent immediately. Save the original packing carton for possible future reshipment.
1.3 Product Dimensions The PVS10005, PVS60085, and PVS60085MR power supply’s dimensions are approximately: 420mm (16.54in) x 88mm (3.46in) x 532mm (20.95in) (WxHxD). These 2U supplies are designed to fit in a standard 19-inch rackmount. Note: All dimensions in the figures below are measured in millimeters (mm). Figure 1.1 – Front Panel Dimensions Figure 1.
1.4 Front Panel Overview 3 2 1 4 5 Figure 1.
1.5 Keypad Overview Figure 1.4 – Keypad Overview Keypad Description Shift key Enables access to secondary functions (Labeled in blue) Meter display button Toggles display between set and measured voltage and current. On/Off Controls the output state. Vset / Slope button Configures output voltage or sets the rise and fall times of the output voltage. Iset / Timer button Configures output current or setup the timer setting. Recall / Save button Saves and recalls instrument settings.
~ Decimal / Local button Inputs a decimal point for values or sets the instrument back to local mode Numeric keypad Enters numeric values for various parameters. Menu button Allows access to the power supply menu settings. Program button Enters program mode settings menu. Enter Confirms setting/parameter changes. Rotary Knob Adjusts numeric settings or selects menu items. Left/Right arrow keys Adjusts cursor position or select menu items.
1.6 Rear Panel Overview 9 1 8 2 3 7 6 4 5 Figure 1.
1.7 Display Overview 1 2 3 4 Figure 1.
2 Getting Started Before connecting and powering up the instrument, please review and go through the instructions in this chapter. 2.
The included AC power cord is safety certified for this instrument operating in rated range. To change a cable or add an extension cable, be sure that it can meet the required power ratings for this instrument. Any misuse with wrong or unsafe cables will void the warranty. SHOCK HAZARD: The power cord provides a chassis ground through a third conductor. Verify that your power outlet is of the three-conductor type with the correct pin connected to earth ground.
Figure 2.1 – AC Power Connection Diagram Refer to the descriptions below to connect the other end of the AC power cord to the AC distribution panel. Do NOT plug the AC power cord into the wall socket prior to connecting ALL three AC power wires to the rear panel and securely mount the safety metal housing over the input receptacle. Doing so may create a shock hazard. Connection of this power supply to an AC power source should be made by a qualified electrician or other qualified personnel.
Figure 2.2 – AC Power Cord Fuse Replacement This power supply does not require a fuse that is user replaceable. There is an internal fuse, in which if blown, may indicate a malfunction in the unit. In this event, contact B&K Precision. Any disassembling of the case or changing the fuse not performed by an authorized service technician will void the warranty of the instrument. 2.
cabinet. Rack mount brackets must be assembled before mounting the unit in a rack. Refer to the following figure to assemble the rack mount brackets. Figure 2.3 – Rack Mount Bracket Assembly 2.3 Output Connections The main DC output terminal is a screw type terminal block located in the rear panel. Due to the high current rating of the power supply, proper wire sizes are necessary for safe connectivity and to prevent wires from overheating.
to the amount of current used for operation: AWG 6 8 10 12 14 16 18 20 22 Imax(A) mΩ/meter 75 1.3 55 2.1 40 3.3 25 5.2 20 8.3 13 13.2 10 21 7 33.5 5 52.8 Table 1 – Wire Gauge Rating Before connecting wires to the output terminals, turn OFF the power supply to avoid damage to the instrument and the device under test (DUT). For safety, load wires must have a wire gauge size large enough to prevent overheating when the power supply operates at maximum short circuit output current.
2.4 Preliminary Check Complete the following steps to verify that the power supply is ready for use. 1. 2. Verify AC input voltage Verify and check to make sure proper AC voltages are available to power the instrument. The AC voltage range must meet the acceptable specification as explained in “2.1 Input Power and Fuse Requirements”. Connect power Connect AC power cord to the AC receptacle in the rear panel and press the power switch to the (ON) position to turn ON the instrument.
Output Check Voltage Check Follow the steps below to check voltage output with no load connected. 1. Turn ON the power supply. The display will show the OFF annunciator above the voltage display. 2. Enable the output by pressing , and the LED next to the button will be lit. The OFF annunciator will change to CV. 3. Using the numeric keypad or the voltage adjust knob and enter a voltage value. The voltage display will now show the value you entered.
2. Short the (+) and (-) output terminals with test leads, shorting bar, or clip. (Refer to “– Wire Gauge Rating” to select appropriate test leads) 3. Using the numeric keypad or the current adjust knob, enter a small current value (i.e. 1.000 A). If entering with numeric keypad, press first, then enter the value and press The current display will now show the value you entered. . 4. Enable the output by pressing , and the LED next to the button will be lit. The OFF annunciator will change to CC. 5.
2. Press the button three times until INFO is blinking. Press , and the display will show the following screen: 3. The model is shown above as PVS10005. 4. Press once again and the firmware version will displayed. 5. The firmware is shown above as 1.52. 6. Press display.
3 Front Panel Operation 3.1 Menu Options All settings and parameters can be configured from the built-in menu system of the power supply. To access the menu, press . The menu system is divided into 5 sections and are organized as follows: SYSTEM COMM MEMORY BEEP DEFAULT PON STAT Select and configure communication interface. Select memory location for save/recall instrument settings. Enable/Disable key sound. Select memory location for save/recall instrument settings. Configure power-on state.
How to Access the Menu Before using the instrument, it is important to be familiarized with its menu structure and learn how to view or change settings and parameters. Follow the steps below to guide you in selecting menu options. 1. From the normal display, press to enter the menu. 2. The selected item will be blinking. Use through the menu selections. 3. When the desired menu section is blinking, press its menu settings. 4. Below is the display when SYSTEM is selected. 5.
8. To exit the menu at any time, press twice. 3.2 Configure Voltage/Current Output Voltage and current can be set from the front panel. Remote sense is also available on the rear panel for voltage compensation at the output. Setting Voltage Follow the steps below to set the output voltage: 1. From the normal front panel display, users can use either the voltage adjust knob or the numeric keypad to enter the setting voltage. 2.
value if the test is needed. Setting Current Follow the steps below to set the output current: 1. From the normal front panel display, users can use either the current adjust knob or the numeric keypad to enter the setting current. 2. If entering using numeric keypad, press first so that the cursor selects the current display. Then, enter the value and press to set the current. 3. To change the cursor position to adjust with the voltage adjust knob, use the keys to move left or right.
Remote Sense Remote sense can be used to compensate for voltage drops up to 6 V (PVS60085/MR) or 10 V (PVS10005) due to resistance from test leads connected to your device under test (DUT), thus providing more accurate output voltage. The power supply is initially set up for local sense mode by default. Refer to the following sections for details of local and remote sense setup. Local Sense By default, the power supply is set up for local sense.
output is connected to the negative end (-) of the load. When this sensing mode is selected the wires connecting between DC outputs to the load must be as short as possible. The local sense is the default configuration with shorting bars connect between (+S) to (+) and (-S) to (-). DO NOT disconnect the wires if remote sense is not used. Doing so will cause erratic behavior and may damage the power supply under certain conditions.
2. Use a small flat blade screwdriver to loosen the wire connection connected between Vo+ and S+ and S- and Vo-. 3. Connect the S+ to the DUT’s positive (+) terminal, and connect the S- to the DUT’s negative (-) terminal. 4. Do not connect any wires to Vo+ and Vo- terminals. 5. Power ON the power supply, and then configure and enable the output. The setup should look like the figure above. DO NOT at any time disconnect the wires from the Vs+ and Vsterminals to the DUT while output is enabled (ON).
The following table shows the various messages that may alert the user: System message IAC TOO HIGH! VBUS NO CHARGE! VAC TOO LOW! VAC TOO HIGH! MEMORY CHECK ERROR! OVER TEMPERATURE PROTECTION! FAN ERROR! EXCEED OVP! EXCEED OCP! CALIBRATION ERROR! AC FREQUENCY ERROR! PFC ERROR! POWER OFF Description AC input current too high Internal PFC VBUS is not charged AC input voltage too low AC input voltage too high Cannot read or write the internal nonvolatile memory Internal temperature is too high and the protect
3.5 SYSTEM Menu All setup procedures and settings explained in this section can be accessed from the SYSTEM menu. To access this menu, press When SYSTEM is blinking, press . . Remote Communication Configuration To set up remote interface connections and settings, refer to chapter “4 Remote Operation”. Save/Recall Instrument Settings The instrument can save up to 100 voltage/current values for quick recall access.
2. Use the rotary knob or the numeric keypad to enter the storage group. Select between 0 – 9. Press 3. To exit the menu at any time, press to save selection. twice. Save Settings 1. Set up output voltage and current settings that you want to save. 2. Then, press and . The display will show the following: 3. Use the current adjust knob or the numeric keypad to enter the memory location in which to store current instrument settings. Select between 0 – 9. Press location.
Instrument settings can only be recalled when the instrument enters this mode. 2. Use the keypad to enter the memory location you want to recall. Enter between 0 – 9. The voltage and current settings of that memory location will be shown on the bottom of the display. 3. Once entered, the saved settings at the location will be immediately recalled. Note: When in Recall mode, users can recall settings from different locations without having to press additional keys each time.
Restore Factory Default Settings All instrument settings can be reset back to their factory default values by doing the following: Note: Restoring the instrument to factory default will change all current instrument settings and parameters back to their default values. 1. From the SYSTEM menu, press button 3 times to select DEFAULT and press . 2. The instrument will return to the normal display and all settings are now restored back to factory default.
CC to CV OFF OFF Voltage 6 V/ms 4 V/ms Current 85 mA/ms 20 mA/ms OFF OFF MASTER MASTER Enable NOT ACTIVE NOT ACTIVE Voltage FRONT FRONT Current FRONT FRONT Program range 5V/5kΩ 5V/5kΩ Monitor range 5V 5V Shut off logic OFF/LOW OFF/LOW Mode OFF OFF VMAX 606.0 V 1010.0 V VMIN 5.0 V 5.0 V IMAX 8.585 A 5.050 A IMIN 0.034 A 0.
GPIB RS-232C Address 1 1 Baud, parity, data, stop 9600, N, 8, 1 9600, N, 8, 1 Baud, parity, data, stop 9600, N, 8, 1 9600, N, 8, 1 1 1 Mode AUTO AUTO IP address 0.0.0.0 0.0.0.0 Mask address 0.0.0.0 0.0.0.0 Gateway address 0.0.0.0 0.0.0.0 RS-485 Address LAN Configure Power-On State The initial Power-On state of the power supply can be configured (voltage, current, output state) by following the steps below: 1.
3. Select one of the settings wanted during power up, and press to save changes. If USER has been selected, set user defined voltage, current and output state. 4. To exit the menu at any time, press twice. 3.6 CONFIG Menu All setup procedures and settings explained in this section can be accessed from the CONFIG menu. To access this menu, press press 1 time and select CONFIG, then press , .
Figure 3.2 – Output Limit Settings Graph Note: At any time during operation, when you are unable to set to a desired voltage or current, check these limit settings to make sure the set value is within the limit’s range. The shaded area illustrated above is the settable range of the output voltage and current. Follow the steps below to configure the output limits: 1. From the CONFIG menu, select LIMIT, then press following screen will display: 34 .
2. Use the keys to select the limit setting value of VMAX, VMIN, IMAX, and IMIN. Press to confirm selection. 3. Use the numerical keys to set the limit value and press confirm the setting. 4. Press to several times to exit the menu setting. Protection Settings Configure Over Voltage Protection (OVP) The PVS overvoltage protection utilizes a hardware comparator that quickly protects the instrument when the voltage presented at the output terminal exceeds the OVP setting voltage.
4. Use numerical keys directly or use adjustment followed by keys with rotary to confirm the OVP value. 5. Press several times to exit the menu setting. 6. If the OVP is on, the OVP LED indicator will be lit up. When OVP protection is tripped during operation, the output will turn off and the following OVP status message will display: To clear the trip status, press once.
4. Use rotary to select the OCP ON or OFF followed by 5. Use numerical keys directly or use adjustment followed by . keys with rotary to confirm the OCP value. 6. Press several times to exit the menu setting. 7. If the OCP is on, the OCP LED indicator will be lit up. When OCP protection is tripped during operation, the output will turn off and the following OCP status message will display: To clear the trip status, press once.
4. Use rotary to select the OPP ON or OFF followed by 5. Use numerical keys directly or use adjustment followed by . keys with rotary to confirm the OPP value. 6. Press several times to exit the menu setting. 7. If the OPP is on, the OPP LED indicator will be lit up. When OPP protection is tripped during operation, the output will turn off and the following OPP status message will display: To clear the trip status, press once.
4. Use rotary to select the OPP ON or OFF followed by 5. Press . several times to exit the menu setting. When CV to CC protection is tripped during operation, the output will turn off and the following CV to CC status message will display: To clear the trip status, press once. Configure CC to CV Protection The PVS CC to CV protection monitors the transition between constant current to constant voltage mode. If this event occurs, the output of the power supply will turn off.
When CC to CV protection is tripped during operation, the output will turn off and the following CC to CV status message will display: To clear the trip status, press once. External Analog Control To control or monitor the output of the power supply using external signals, refer to the following figure of the DB25 connector located in L is a TTL compatible signal. The the rear panel. The logical signal ○ A cannot exceed the range of 0 to 12VDC.
Figure 3.
9 Voltage Program ○ 10 Current Program ○ 11 Voltage Monitor ○ 12 Ground 13 CV/CC A A A L ○ I Input 0-5 V / 0-5 kΩ or 0-10 V / 0-10 kΩ for voltage output setting, full scale input equals maximum output voltage I Input 0-5 V / 0-5 kΩ or 0-10 V / 0-10 kΩ for current output setting, full scale input equals maximum output current O Output 0-5 V / 0-10 V represents power supply output voltage, full scale output equals maximum output voltage High: constant voltage O Low: constant current 14
21 Local/Analog State 22 Ground 23 Ground 24 Current Monitor L ○ High: front panel controlled O Low: rear analog controlled ○ A O Output 0-5 V / 0-10 V represents power supply output current, full scale input equals maximum output current Enabling/Disabling External Control Follow the steps below to configure and enable the external control interface. 1. Press the button and press the CONFIG is blinking and press 2. Press button one time until .
Enable+ (Pin 1) and Enable- (Pin 14) are opened, the button is disabled and the power supply’s output will remain OFF. Pressing the button will not turn ON the output, and the display will also show ENA. When Enable+ (Pin 1) and Enable- (Pin 14) are shorted, the is enabled. This will enable the button button to allow front panel control of the output state. If the ENA is shown on the LCD screen, press once so that it disappears before attempting to press the button.
ACTIVE BMODE Opened Disable OFF --- Shorted Disable ON --- Local/Analog Control Pin 8 can be used to select the control mode (local or analog) of the power supply’s output. When the input command for this pin is at high level (or open), the control mode will be local. When input command for this pin is at low level (or shorted to GND), the control mode will be analog.
Power OK Signal Pin 16 is used to indicate whether a fault condition is present in the power supply. Normally this pin will output a logic high (5V). When a fault occurs, this pin will output a logic low (0V). Fault conditions are defined as follows: 1. Overvoltage protection 2. Overcurrent protection 3. Overpower protection 4. CV to CC protection 5. CC to CV protection 6. Overtemperature protection 7. AC line failure 8. Enable signal 9. Shutoff signal 10. Remote failure 11.
2. Press press button three times until EXTCTRL is blinking and . 3. Press button one time to select VOLTAGE and press . 4. Use the rotary knob to select EXT-V (external voltage) or EXT-R (external resistance) option and press control, select FRONT. 5. Use the . To have front panel keys to select PROG RNG (program range) and press . 6. Now use the rotary knob to select between 10V/10K or 5V/5K and press 7. Press . several times to exit the menu setting.
value by connecting a resistance value of 0-5 kΩ (0-5 V / 0-5 kΩ mode) or 0-10 kΩ (0-10 V / 0-10 kΩ mode) between Pin 9 and Pin 22 as shown below. 0 – 5 kΩ or 0 – 10 kΩ Figure 3.5 - Analog Voltage Programming Diagram (Resistor Mode) Current Program This function is able to program the current output by connecting an external DC voltage (voltage mode) or an external resistor (resistor mode) to Pin 10. For this function to be enabled, the output control must be in analog mode.
6. Now use the rotary knob to select between 10V/10K or 5V/5K and press 7. Press . several times to exit the menu setting. Voltage Mode Under voltage mode, the user may control the full scale current output value through Pin 10, by inputting a voltage level of 0-5V (0-5 V / 0-5 kΩ mode) or 0-10V (0-10 V / 0-10 kΩ mode) as shown below. 0 – 5V or 0 – 10V Figure 3.
Voltage Monitor This function is able to monitor the voltage output using Pin 11 and one of the ground pins (i.e. Pin 22), which can be connected to a digital voltage meter (DVM) or other voltage monitoring device, as shown below. The output control must be in analog mode to use this function. The monitoring of the output voltage range (which reflects 0 to full scale of the power supply’s output voltage) can be selected between 0-5 V and 0-10 V. Figure 3.
of the ground pins (i.e. Pin 23), which can be connected to a digital voltage meter (DVM) or other voltage monitoring device, as shown below. The output control must be in analog mode to use this function. The monitoring of the output voltage range (which reflects 0 to full scale of the power supply’s output voltage) can be selected between 0-5 V and 0-10 V. Figure 3.9 - Analog Current Monitor Diagram Follow the steps below to configure the voltage program. 1.
Follow the steps below to configure the shut off control. 1. Press the button and press the CONFIG is blinking and press 2. Press press button one time until . button three times until EXTCTRL is blinking and . 3. Press button five times to select SHUT OFF and Press The following will be on the display: 4. Use the rotary knob to select the setting of OFF/Low (Falling Edge) or ON/Low (Rising Edge) and press 5. Press . several times to exit the menu setting.
Note: Shut off can only occur when the power supply receives an edge trigger. Maintaining that pin constantly at a high level or low level will not trigger a shut off.
Figure 3.10 – Parallel Connection Diagram After wiring is complete, configure one of the supplies as the Master and the other three as Slave A, B, and C. After one of the supplies is configured to be the Master, it will start searching for all Slaves that are connected to the Master. To operate correctly, the user must set up the Slaves before the Master. Follow the steps below to set a supply to master or slave mode. 1. Press the button and press the CONFIG is blinking and press 54 .
2. Press press button four times until PARALLEL is blinking and . 3. Press to select the parallel MODE. 4. Use the rotary knob to select the setting of OFF or ON and press . 5. Press button one time until ROLE is blinking and press 6. Use the rotary knob to set the master or slave and press . Use the same procedure to set up MASTER, SLAVE A, SLAVE B, or SLAVE C for the other power supplies. 7. Press several times to exit the menu setting.
In parallel mode, the slave unit is remote controlled by the master unit. The keypad will be locked. To exit the parallel mode, press key to access the parallel mode ON/OFF setting in the menu. Turn off to disable parallel mode control. Master Unit After all of the slave units have been set, the master unit can be set by the same procedure with the role set to MASTER. As shown below, the master unit will start to search for the slave units.
While in parallel connection mode, the output voltage of each power supply should be set to equal value. If the voltage value of each unit is not the same, the higher output voltage will feed back to the lower voltage unit and damage its internal components. Do not connect multiple power supplies in series as it may cause damage or malfunction. 3.7 PV Simulation Typically, a solar array is connected to an inverter, which converts the panel’s output from DC to AC.
effectiveness of the inverter’s MPPT algorithm. Figure 3.11 – PV Simulation Curve Solar arrays consist of multiple solar cells characterized by a complex voltage and current profile that is represented in an I-V curve.
the I-V profile. b) Manually generate the I-V curve by creating a user-defined table of points in the SAS software. Up to 1024 I-V points can be downloaded to the PVS supply’s non-volatile memory. Once one of these I-V curves has been selected and set from the front panel, the power supply will operate in PV mode and output a voltage and current value according to the active I-V profile and load conditions.
3. Use the keys to move the cursor to the CURVE field and use the rotary knob to select the desired curve number. The curve number can be 1-16 or TABLE (table of up to 1024 I-V points.) 4. Once again, use the keys to move the cursor to the EDIT field and press . The following display will be shown: 5. Use the keys to move the cursor through the four parameter settings. Use the numerical keys to enter parameter values and press settings.
illustration of how programs are structured, stored, and recalled. … P3 P9 P2 P1 Repeat Program Figure 3.12 – Program Structure Illustration Next Program (P3) Configure Program Parameters Actions The program function is used to edit a test pattern composed by steps. Each step contains the voltage, current, time, and action. There are three types of actions that can be set: ON, OFF, and END. The ON action enables the current step and will be executed during operation.
Options Users also have the ability to utilize the REPEAT and NEXT PROG options. The REPEAT option allows the user to execute the same program as many times, continuously, as needed. The NEXT PROG option allows the user to select the next program to execute once the current program ends. Please be aware that REPEAT and NEXT PROG options are part of a program, so if one program’s NEXT PROG option is pointed to a previous program, the power supply will cycle these programs infinitely.
2. Use the rotary knob to select the setting of OFF or ON. 3. Use the buttons to move the cursor to the PROGRAM selection and use the rotary knob to select a program between 1-9. 4. Use the selection and press buttons to move the cursor to the EDIT . The following screen will be displayed: 5. As described previously, each step has the following parameters to be edited: step number, action, voltage, current, and time.
7. At this point, the user may select how many times they wish the current program to repeat in the REPEAT field, by using the keypad or the rotary knob. 8. Press to select NEXT PROG parameter and use the rotary knob to select a value of NONE or a value of 1-9. As described previously, the user has the ability to queue another program once the current program is finished. By default, NONE is placed as the parameter value. 9. Press to complete the program configuration. 10.
example, if the xx is 01 and yy is 03, this means that there are three repeats of this program and it is now running the first repeat. 3.9 Timer Function The timer function is a countdown clock, which allows the user to setup how long the output will be enabled when it is initially turned ON. When it is running, a clock on the display counts down until it reaches zero and the output will turn OFF. An * indicator on the display will flash to indicate the timer is running.
3. Use the buttons to move the cursor to the ON/OFF parameter and use the rotary knob to change ON or OFF. 4. Press to complete the timer setting. 5. Pressing the key will enable the output and start the timer. 3.10 Slew Rate Setting The power supply has the capability of controlling the output voltage and current slew rate. The timing can be configured for the rising/falling edge between voltage and current output transitions. Follow the steps below to configure the voltage or current slope . 1.
4 Remote Operation There are several interfaces available for remote communication: USB, RS-232, GPIB, Ethernet, and RS-485. With all these interfaces, this power supply is very flexible to be controlled remotely. Users can program the power supply by using the SCPI (Standard Commands for Programmable Instruments) commands over any one of the remote interfaces. This section will describe how to set up all of the supported interfaces.
1. From the SYSTEM menu, select Communication and press 2. Press . button three times until RS-232 is blinking and press to configure the settings for RS-232 remote communication. The following display will be shown: 3. Use keys to select between each serial settings and press to configure the corresponding settings. 4. To change the baud rate or any of the parameter settings, use the rotary dial. 5. Press to save each setting and the display will return to the RS232 menu. 6.
Therefore, a delay between commands to let the power supply have enough time to process them is mandatory. 7. All serial settings must match the settings configured on the PC in order for communication to link successfully. GPIB Each model can be configured with a GPIB address from 1-30. Follow the instructions below to select and configure the GPIB interface for remote operation. 1. From the SYSTEM menu, select Communication and press . 2.
Follow the instructions below to select and configure the RS-485 interface for remote operation. PC Figure 4.
1. From the SYSTEM menu, browse and select Communication and press 2. Press . button two times until RS-485 is blinking and press to configure the settings for RS-485 remote communication. The following display will be shown: 3. Use keys to select between each settings and press to configure the corresponding setting. 4. To change the parameter settings, use the rotary dial. 5. Press to save each setting and the display will return to the RS485 menu. 6.
more supplies have the same address, the RS485 protocols will create communication errors. 8. Repeat the above steps for each power supply that you want to connect together and control, making sure that each of them have a different address assigned. 9. Connect the USB of the first supply (address 1) to the computer, then use the Multi-Unit Programming commands to communicate with each power supply. Refer to “4.7 Multi-Unit Programming Commands”.
5. Press to save each setting and the display will return to the LAN menu. 6. The following lists the options that can be changed for each setting: Mode: Web, Tel, Socket IP Addr: 000.000.000.000 Gway: 000.000.000.000 Mask: 000.000.000.000 Web server There is an embedded web server GUI that can access the power supply via LAN interface using a web browser.
Figure 4.2 – Web Server Login Page 6. A password is required to login and access the menu items on the page. The default admin password 123456. The web server menu items are described below: HOME The HOME page provides general information of the power supply: Model, Manufacturer, Firmware version, MAC address, IP address, RS-232C setting, and RS-485 setting.
Figure 4.3 – Web Server Home Page CONFIG The CONFIG page provides the protection settings (OVP, OCP, OPP, CV to CC, and CC to CV) and output related parameters such as voltage/current slope and limit settings.
Figure 4.4 – Web Server Configuration Page CONTROL The CONTROL page provides the general control of the power supply such as output On/Off setting as well as the voltage and current setting. The command line for SCPI commands can also be accessed here.
Figure 4.5 – Web Server Control Page The Log Out will exit the web page and go back to login screen. Telnet connection The power supply can be connected via LAN (Ethernet) interface using Telnet client with socket port 23 or 5024. Socket connection Socket connection is available for communication via LAN (Ethernet) interface. The instrument uses the TCP/IP protocol for communication. Users can use socket port 5025 to open a raw socket connection for sending remote commands.
4.2 Parameter Definition The PVS power supply supports communication protocols, which include standard SCPI commands and a few proprietary commands that follow the SCPI convention. The SCPI interface enables users to operate the power supply through a computer or a terminal equipped with IEEE488.2 GPIB, RS-232, or USB interface. SCPI IEEE-488.2 also supports multi-unit control allowing a user to control up to 30 power supplies. The following table lists all of the numerical parameters.
clear all errors/events from the memory, the “*CLS” command can be used. The following table lists all the known errors that can be encountered. Error -000 -102 -103 -108 -109 -113 -131 -138 -200 Description No error Syntax error Invalid separator Parameter not allowed Missing Parameter Undefined header Invalid suffix Suffix not allowed Execution error 4.4 SCPI Common Commands The power supply is compatible with all IEEE-488.2 and SCPI commands.
*RCL Recall setting from memory *RST Reset *SAV Save setting to memory 4.5 SCPI Common Subsystem Subsystem commands are specific to functions. They can be a single command or a group of commands. The groups are comprised of commands that extend one or more levels below the root. The subsystem commands are arranged alphabetically according to the function they perform.
EXTV/1, EXTR/2 | ?> :MONitor :RANGe < 5V/0, 10V/1 |?> :PROGram :RANGe <5V | 5K/0, 10V | 10K/1 | ?> :SHUToff :VOLTage :MODE :PARAllel [:STATe] :ROLE < MASTer/0, SLAVEA/1, SLAVEB/2, SLAVEC/3,| ?> :CHAIn [:STATe] :ADDress FETCh :CURRent? :VOLTage? MEASure :ALL? :CURRent? mode Set or return external monitor range Set or return external program range Set or return shut-off logic state Set or return voltage external control mode Set or
:VOLTage? MEMory [:NUMber] :ISET :SAVE :VSET OUTput [:STATe] :LIMit :CURRent :VOLTage :PROtection :CLEar :SLOPe :CURRent :VOLTage :MODE? PROGram [:NUMber] :CLEar :ALL :NEXT :REPeat :RUN :STATe :STEP Return the measured output voltage Memory subsystem Select or return memory number, range 1-9 Set or return current value Save memory subsystem parameters Set or return voltage va
[:NUMber] :ACTion :CURRent :ONTime :VOLTage :TOTAl? PROTection [:STATe] ? :CCCV :CLEar :CVCC :OCP [:STATe] :LEVel Set or return step number Set or return step n action state Set or return step n current value Set or return step n time value Set or return step n voltage value Return program total n step numbers Protection subsystem Return the protected state, 0=no protection happened, 1=OVP,
PVSIMulation :CURVe [:NUMber] :PARAMeter :SELEct :POWer [:STATe] :TABLe [:NUMber] :PARAMeter :START :STATe? :STOP [SOURce] :CURRent [:LEVel] :PROTection [:STATe] level PV Simulation subsystem Set the curve number n to write the parameters to Set or return the parameters (Isc, Voc, Imp and Vmp) of the curve number n Set or return the selected curve to be running Set or return the simulated output power fr
:LEVel :VOLTage [:LEVel] :PROTection [:STATe] :LEVel SYStem :BEEP :ERRor? :GPIB :ADDRess :GROUp :IP :ADDRess :CONFig :GATEway :MASK
:PMAX? :POWer :CURRent :STATe :TYPE :VOLTage :RECall :DEFault :RS232 :BAUDrate :DATAbit :STOPbit :RS485 :BAUDrate :DATAbit :STOPbit :SERial? TIMER [:STATe] :HOUR :MINute :SECond Return the power supply’s maximum output power Power on state Set or return user define current value Set or return user define output state Set or return power on m
4.
4.7 Multi-Unit Programming Commands The multi-unit programming commands used by the power supply include a carriage return (CR) character for termination of all ASCII strings. For all configuration commands (except for commands listed under Synchronous Control Commands section), the instrument will return a string “OK” to indicate command sent successfully. If an error occurs, it will return an error. See the “Error list” section for details.
before using any of these commands to control that unit’s output.
connected in the RS-485 chain at once. Note that these commands will not return an “OK” string upon making a configuration. Command GRST GCLS GCLR GPV GPC GOUT GOV GOVP GOC GOCP Description Reset command.
Status Define The “CST?” command queries the device status.
For example, if the OVP, OCP, and OPP ON/OFF switches are ON and the output is ON, the return value will be E40000. Byte0 Byte1 Byte2 Bit7 1 0 0 Bit6 1 0 0 Bit5 1 0 0 Bit4 0 0 0 Bit3 0 0 0 Bit2 1 0 0 Bit1 0 0 0 Bit0 0 0 0 String E4 00 00 5 Calibration B&K Precision recommends a calibration interval of one year to ensure that the power supply meets specifications. This instrument features closed-case calibration.
2. Enter the password 13579, using the numeric keypad, and press the key to confirm. The following display will be shown: Below is a list of calibration options: 1. 2. 3. 4. 5. 6. 7. 8. Output voltage Output current Overvoltage protection Overcurrent protection External voltage programming External current programming Constant current of external voltage programming Constant current of external current programming These options are only accessible in the calibration menu.
Figure 5.1 – Voltage Calibration Diagram 2. Set the DMM to a DC voltage measurement. Select the VOLTAGE parameter in the CAL menu and press following will be displayed: . The The P: 1/4 signifies there are 4 points to be calibrated, and the current calibration point is 1. 3. Input the voltage reading from the DMM in the METER parameter by using the numeric keypad and press . 4. Repeat this procedure by entering the meter readings for points 2 – 4. 5.
Follow the steps below to perform the current calibration: 1. Connect the DMM to the output of the power supply as shown in the figure below: Figure 5.2 – Current Calibration Diagram 2. Set the DMM to a DC current measurement. Select the CURRENT parameter in the CAL menu and press following will be displayed: . The The P: 1/4 signifies there are 4 points to be calibrated, and the current calibration point is 1. 3.
4. Repeat this procedure by entering the meter readings for points 2 – 4. 5.3 OVP Calibration The overvoltage protection (OVP) calibration can be executed right after the voltage calibration or the user can select OVP in the CAL menu. Follow the steps below to perform the overvoltage protection calibration: 1. Disconnect any load from the output terminal of the power supply. 2. Select the OVP parameter in the CAL menu and press following will be displayed: . The 3.
5.4 OCP Calibration The overcurrent protection (OCP) can be executed right after the current calibration or the user can select OCP in the CAL menu. Follow the steps below to perform the voltage calibration: 1. Short the output + and - terminals of the power supply as shown below: Figure 5.3 – OCP Calibration Diagram 2. Select the OCP parameter in the CAL menu and press following will be displayed: . The 3. Press to run the OCP calibration or press to abort.
5.5 External Voltage Programming Calibration Follow the steps below to perform the external voltage programming calibration: 1. Connect the DMM to the external analog control terminals of DB25 connector (+ on the DMM to Pins 9 and 11, - on the DMM to Pin 22 (GND)) on the rear of the power supply, as shown in the figure below: Figure 5.4 – External Voltage Programming Calibration Diagram 2. Set the DMM to a DC voltage measurement.
The P: 1/2 signifies there are 2 points to be calibrated and the current calibration point is 1. 3. Input the voltage reading from the DMM in the METER parameter by using the numeric keypad and press . 4. Repeat this procedure by entering the meter readings for points 2. 5.6 External Current Programming Calibration Follow the steps below to perform the external current programming calibration: 1.
2. Set the DMM to a DC voltage measurement. Select the EXT_CURR parameter in the CAL menu and press following will be displayed: . The The P: 1/2 signifies there are 2 points to be calibrated and the current calibration point is 1. 3. Input the voltage reading from the DMM in the METER parameter by using the numeric keypad and press . 4. Repeat this procedure by entering the meter readings for points 2. 5.
Figure 5.6 – CC Calibration of External Voltage Diagram 2. Set the DMM to a DC current measurement. Select the EXT_V_CC parameter in the CAL menu and press following will be displayed: . The 3. Input the current reading from the DMM into the METER parameter by using the numeric keypad and press . 5.8 CC Calibration of External Current Follow the steps below to perform the calibration procedure for the constant current source of the external current programming by resistor mode: 1.
Figure 5.7 – CC Calibration of External Current Diagram 2. Set the DMM to a DC current measurement. Select the EXT_I_CC parameter in the CAL menu and press be displayed: . The following will 3. Input the current reading from the DMM into the METER parameter by using the numeric keypad and press 102 .
6 Troubleshooting Guide Below are some frequently asked questions and answers. Please check if any apply to your power supply before contacting B&K Precision. General Q: I cannot power up the power supply. - Check that the power cord is securely connected to the AC input and there is live power from your electrical AC outlet. - Verify that the AC power coming from the mains contains the correct voltage. The power supply can accept a specific range of AC input voltages. Refer to section 2.1.
Remote Control Q: I am trying to send the commands over USB/RS232, but it does not seem to respond. - Check that you are sending ASCII strings that are terminated with a CR (carriage return) and LF (linefeed) character. - For RS-232 and RS-485, check that the baud rate, parity, data bits, stop bit, and flow control settings match with the settings configured on the software interface. - For RS-485, check that the wires are correctly daisy chained between multiple units.
7 Specifications Note: All specifications apply to the unit after a temperature stabilization time of 15 minutes over an ambient temperature range of 23 °C ± 5 °C. Specifications are subject to change without notice. Environmental Conditions: This power supply is designed for indoor use and operated with maximum relative humidity of 90%. Model PVS60085 PVS60085MR Output Rating Output Voltage1 0 – 600 V Output Current2 0 – 8.5 A Output Power 5100 W 3000 W Line Regulation Voltage 60 mV Current 4.
Programming Accuracy Voltage 400 mV 700 mV 3 Current 0.03% + 3.5 mA 0.03% + 2 mA (± (%output +offset)) Readback Accuracy ± (%output +offset) ( Voltage 0.05 % + 300 mV 0.05 % + 500 mV 3 Current 0.1 % + 8.5 mA 0.1 % + 5 mA General Transient Response4 ≤ 0.5 ms ≤ 0.5 ms TimeAdjustment Range OVP 5 – 660 V 5 – 1100 V OVP Accuracy 6V 10 V OCP Accuracy 85 mA 50 mA Command Response 20 ms 20 ms 5 Time Efficiency 90 % 90 % Power Factor 0.99 0.
1 Minimum voltage is guaranteed to 5V. current is guaranteed to maximum 0.4% of rated output current. 3 Current accuracy is applied when output power > 0.1% of full power. 4 Time for output voltage to recover within 0.5% of its rated output for a load change 50 – 100% of rated output current. 5 90% typical at nominal line and max load. 6 10% of output de-rated when operating below 190 VAC. 2 Minimum To ensure the most current version of this manual, please download the latest version here: http://www.
SERVICE INFORMATION Warranty Service: Please go to the support and service section on our website at www.bkprecision.com to obtain a RMA #. Return the product in the original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Non-Warranty Service: Please go to the support and service section on our website at www.bkprecision.com to obtain a RMA #.
LIMITED THREE YEAR WARRANTY B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of three years from date of purchase. B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt.
22820 Savi Ranch Parkway Yorba Linda, CA 92887 www.bkprecision.com © 2014 B&K Precision Corp.