User Manual
103
Figure 5.7 – CC Calibration of External Current Diagram
2. Set the DMM to a DC current measurement. Select the EXT_I_CC
parameter in the CAL menu and press . The following will
be displayed:
3. Input the current reading from the DMM into the METER
parameter by using the numeric keypad and press .