User Manual
are sent too fast to the power supply, the internal
buffer may overrun and cause a communication
error.
Therefore, a delay between commands to let the power supply have enough time to process them is
mandatory.
All serial settings must match the settings configured on the PC in order for communication to link
successfully.
4.1.5 GPIB
Each model can be configured with a GPIB address from 1-30. Follow the instructions below to
select and configure the GPIB interface for
remote operation.
1.
From the
menu, select
Communication
and press
.
2.
When
GPIB
is select, press
to configure the settings for GPIB remote communication. The
following display will be shown:
3.
Press
to configure the GPIB Address to which the power
supply will be assigned to. Use the
rotary dial or numeric keypad
to enter a value from 1-30.
4.
Press
to save each setting and the display will return to the GPIB menu.
5.
Make sure the address assigned in the power supply unit
matches the address the data commands
are being sent to.