User Manual

To clear the trip status, press
once.
3.4.3.1 Configure CV to CC Protection
The MR CV to CC protection monitors the transition between constant
voltage to constant
current mode. If this event occurs, the output of the power supply will turn off.
Follow the steps below to set the CV to CC Protection:
1.
Press
button
,
select
CONFIGURATION
, then press
.
2.
P
ress
button one time until
PROTECT
is select and press
. The display will then show
OVP.
3.
Press
button three times until CV -> CC is displayed and press .
4.
Use rotary to select the CV->CC ON or OFF followed by
.
5.
Press
several times to exit the menu setting.
When CV to CC protection is tripped during operation, the output will turn off and the following CV to CC
status message will display:
To clear the trip status, press
once.
3.4.3.2 Configure CC to CV Protection
The MR CC to CV protection monitors the transition between constant
current to constant
voltage mode. If this event occurs, the output of the power supply will turn off.
Follow the steps below to set the CV to CC Protection:
1.
Press
button
,
select
CONFIGURATION
, then press
.
2.
Press
button one time until
PROTECT
is select and press
. The display will then show
OVP.
3.
Press
button four times until CC->CV is displayed and press .