User Manual

2
3 Use the keys to select the limit setting value of V max,
V min, I max, and I min. Press
to confirm selection.
4
Use the numerical keys to set the limit value and press
to confirm the setting.
5
Press
several times to exit the menu setting.
3.4.2 Protection Settings
3.4.2.1 Configure Over Voltage Protection (OVP)
The MR over voltage protection utilizes a hardware comparator that
quickly protects the
instrument when the voltage presented at the output terminal exceeds the OVP setting voltage.
Follow the steps below to set the OVP limit:
1.
Press
button
,
select
CONFIGURATION
, then press
.
2.
Press
button one time until
PROTECT
is select and press
. The display will then show
OVP,
press
.
3.
Use numerical keys directly or use
keys with rotary
adjustment followed by
to
confirm the OVP value.
4.
Press
several times to exit the menu setting.
When OVP protection is tripped during operation, the output will turn off and the following OVP status
message will display:
To clear the trip status, press
once.
3.4.2.2 Configure Over Current Protection (OCP)
The MR overcurrent protection utilizes a hardware comparator that
quickly protects the
instrument when the current presented at the output terminal exceeds the OCP setting current.
Follow the steps below to set the OCP limit:
1.
Press
button
,
select
CONFIGURATION
, then press
.