5 Calibration
B&K Precision recommends a calibration interval of one year to ensure that the power supply
meets specifications. This instrument features closed-case calibration. To perform the
calibration, the following equipment is required:
5 ½ digit digital multimeter (DMM); B&K Precision 5491B or equivalent
The following calibration instructions may be used by authorized technicians or
calibration personnel only. If you are not authorized, do not attempt to calibrate
the instrument yourself, as it may damage the instrument and void the warranty.
To enter the calibration procedure, follow the steps below:
1.
Press the
button and press the
button until SYSTEM is selected and press
. Keep
press the
button until CALIBRATION is selected and press
.
2.
Enter the password
13579
, using the numeric keypad, and press
the key to confirm. The
following display will be shown:
Below is a list of calibration options:
VOLTAGE
CURRENT
OVP
OCP
EXT VOLTAGE
EXT CURRENT
EXT VOLTAGE CC
EXT CURRENT CC