[:LEVel]
:RANGe <NRf> Set constant current range. MAX Y
:RANGe? Query constant current range. Y
:SLEW Y
[:BOTH] <NRf>
Set both positive and negative slew rate for
current
MAX Y
:POSitive <NRf> Set positive slew rate for current MAX Y
:POSitive? Query positive slew rate for current Y
:NEGative <NRf> Set negative slew rate for current Y
:NEGative? Query negative slew rate for current Y
:PROTection
[:STATe] <b> Disable or enable current protection ON
[:STATe]? Query current protection state
:LEVel <NRf> Set current protection level MAX
:LEVel? Query current protection level
:DELay <NRf> Set current protection delay time 3
:DELay? Query current protection delay time
:TRANsient
:MODE <name>
Set current transiton mode (CONTinuous,
PULSe, or TOGGle)
CONTinuous
:MODE?
Query current transition mode
:ALEVel <NRf> Set current transition level A MAX
:ALEVel? Query current transition level A
:AWIDth <NRf> Set current transition width A 500uS
:AWIDth? Query current transiton width A
:BLEVel <NRf> Set current transition level B MIN
:BLEVel?
Query current transition level B
:BWIDth <NRf> Set current transition width B 500uS
:BWIDth? Query current transiton width B
:HIGH <NRf> Set the voltage specification of upper limit