Owner manual

76
Channel 18-11 represents channels 8 to 1 for MDL002
extended mainframe.
1 channel will be shorted.
0 channel will not be shorted.
Bit
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Channel
18
17
16
15
14
13
12
11
8
7
6
5
4
3
2
1
Examples
PROG:SEQ:SHOR 683
683 = 1000100
Sets channel 3 and 7 for short testing.
Query Syntax
PROG:SEQ:SHOR? <n>
Returned Parameters
<NRf+>
:PROGram:SEQuence:ONTime
The command is used to specify the input ON time for the selected sequence, which controls how long
to enable the inputs for the selected sequence.
Command Syntax
PROGram:SEQuence:ONTime <n>, <NRf>
Parameters
<n> - Sequence number, 1 to 10.
<NRf> - On time, 0.01 to 60.0 seconds.
Examples
PROG:SEQ:ONT 4, 5
Sets sequence 4 ON time to 5 seconds.
Query Syntax
PROG:SEQ:ONT? <n>
Returned Parameters
<NR1>
:PROGram:SEQuence:OFFTime
The command is used to specify the input OFF time for the selected sequence, which controls how long
to disable the inputs for the selected sequence.
Command Syntax
PROGram:SEQuence:OFFTime <n>, <NRf>
Parameters
<n> - Sequence number, 1 to 10.
<NRf> - Off time, 0.01 to 60.0 seconds.
Examples
PROG:SEQ:OFFT 4, 5
Sets sequence 4 OFF time to 5 seconds.
Query Syntax
PROG:SEQ:OFFT? <n>
Returned Parameters
<NR1>