This command is used to control auto ranging for voltage. With auto ranging enabled, the instrument
automatically goes to the most sensitive range to perform the measurement.
The auto range command (:RANGe:AUTO) is coupled to the command that manually selects the
measurement range (:RANGe <n>). When auto range is enabled, the parameter value for :RANGe <n>
changes to the automatically selected range value. Thus, when auto range is disabled, the instrument
remains at the automatically selected range. When a valid:RANGe <n> command is sent, auto ranging
disables.
Command Syntax
SENSe:VOLTage[:DC]:RANGe:AUTO <bool>
Parameters
0 | 1 | OFF | ON
*RST Value
ON
Examples
SENS:VOLT:RANG:AUTO 1
Query Syntax
SENSe:VOLTage[: DC]:RANGe:AUTO?
Returned Parameters
0 | 1
Related Commands
SENS:VOLT:RANG
Program Commands
The program subsystem is used to control the parameters of the built-in automatic test function.
:PROGram:ACTive:CHANnel
The command is used to specify the active channels for the automatic test function. When a channel is
active, some parameters for the program setup and selec the channel for the test.
Command Syntax
PROGram:ACTive:CHANnel <NRf+>
Parameters
0 – 65535
The value, in decimal, is based on the binary
representation of the active channels which are
represented by the bits, according to the following table.
Note: Channel 18-11 represents channels 8 to 1 for
MDL002 extended mainframe.
1 – channel is active.
0 – channel is not active.
Bit
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Channel
18
17
16
15
14
13
12
11
8
7
6
5
4
3
2
1
Examples
PROG:ACT:CHAN 683
683 = 1000100
Sets channel 3 and 7 active.