Owner manual
67
This command is used to set series resistance when the load is in CZ mode.
Command Syntax
[SOURce:]IMPedance:RESistance <NRf+>
Parameters
0 through 7500 ohms
Unit
R (ohms)
*RST Value
0
Examples
IMP:RES 5
Query Syntax
[SOURce:]IMPedance:RESistance?
Returned Parameters
<NR3>
[SOURce:]IMPedance:INDuction
Channel Specific
This command is used to set the series inductance when the load is in CZ mode.
Command Syntax
[SOURce:]IMPedance:INDuction <NRf+>
Parameters
0 through 65000uH
Unit
uH
*RST Value
0
Examples
IMP:IND 5
Query Syntax
[SOURce:]IMPedance:INDuction?
Returned Parameters
<NR3>
[SOURce:]IMPedance:CAPacitance
Channel Specific
This command is used to set the parallel capacitance value when the load is in CZ mode.
Command Syntax
[SOURce:]IMPedance:CAPacitance <NRf+>
Parameters
20 through 65000uF
Unit
uF
Examples
IMP:CAP 5
Query Syntax
[SOURce:]IMPedance:CAPacitance?
Returned Parameters
<NR3>
[SOURce:]IMPedance:HIGH
[SOURce:]IMPedance:LOW
This command is used to set the high and low voltage determined when the load is in CZ mode.