Owner manual

64
Command Syntax
[SOURce:]POWer[:LEVel][:IMMediate] <NRf+>
Parameters
0 to max. rated power
Unit
W (power)
*RST Value
MINimum
Examples
POW 5 POW:LEV 3.5
Query Syntax
[SOURce:]POWer[:LEVel][:IMMediate]?
Returned Parameters
<NR3>
Related Commands
POW:RANG
[SOURce:]POWer:RANGe
Channel Specific
This command sets the power range of the load. This limits the settable range for power.
Command Syntax
[SOURce:]POWer:RANGe <NRf+>
Parameters
0 to max. rated power.
Unit
W (power)
Examples
POW:RANG 15 SOUR:POW:RANGE MIN
Query Syntax
[SOURce:]POWer:RANGe?
Returned Parameters
<NR3>
[SOURce:]POWer:HIGH
[SOURce:]POWer:LOW
Channel Specific
This command sets the voltage high and low limit determined when in constant power mode.
Command Syntax
[SOURce:]POWer:HIGH <NRf+>
[SOURce:]POWer:LOW <NRf+>
Parameters
0 to max. rated voltage
Unit
V (volts)
Examples
POW:HIGH 5
Query Syntax
[SOURce:]POWer:HIGH? [SOURce:]POWer:LOW?
Returned Parameters
<NR3>