Owner manual
63
[SOURce:]RESistance:TRANsient:ALEVel
[SOURce:]RESistance:TRANsient:BLEVel
Channel Specific
This command specifies the transient level of the input resistance. The transient function switches
between level A and level B.
Command Syntax
[SOURce:]RESistance:TRANsient:ALEVel <NRf+>
[SOURce:]RESistance:TRANsient:BLEVel <NRf+>
Parameters
0 to max. rated resistance.
Unit
Ω (ohms)
Examples
RES:TRAN:ALEV 5 RES:TRAN:BLEV 0.5
Query Syntax
[SOURce:]RESistance:TRANsient:ALEVel?
[SOURce:]RESistance:TRANsient:BLEVel?
Returned Parameters
<NR3>
Related Commands
RES
[SOURce:]RESistance:TRANsient:AWIDth
[SOURce:]RESistance:TRANsient:BWIDth
Channel Specific
This command specifies the transient pulse width of the input resistance.
Command Syntax
[SOURce:]RESistance:TRANsient:AWIDth <NRf+>
[SOURce:]RESistance:TRANsient:BWIDth <NRf+>
Parameters
1 to 65535 us
Unit
S (second)
*RST Value
500 uS
Examples
RES:TRAN:AWID 0.001 RES:TRAN:BWID 0.02
Query Syntax
[SOURce:]RESistance:TRANsient:AWIDth?
[SOURce:]RESistance:TRANsient:BWIDth?
Returned Parameters
<NR3>
Related Commands
RES
[SOURce:]POWer
Channel Specific
This command sets the power of the load when operating in constant power mode.