[SOURce:]RESistance:HIGH
[SOURce:]RESistance:LOW
Channel Specific
This command is used to set the voltage high and low limit determined when the load is in constant
resistance mode.
Command Syntax
[SOURce:]RESistance:HIGH <NRf+>
[SOURce:]RESistance:LOW <NRf+>
Parameters
0 to max. rated voltage
Unit
V (volts)
Examples
RES:HIGH 5
Query Syntax
[SOURce:]RESistance:HIGH?
[SOURce:]RESistance:LOW?
Returned Parameters
<NR3>
[SOURce:]RESistance:TRANsient:MODE
Channel Specific
This command selects the operating mode of the transient generator as follows in constant resistance
mode.
CONTinuous The transient generator puts out a continuous pulse stream upon receipt of a trigger.
PULSe The transient generator puts out a single pulse upon receipt of a trigger.
TOGGle The transient generator toggles between two levels upon receipt of a trigger.
Command Syntax
[SOURce:]RESistance:TRANsient:MODE <mode>
Parameters
CONTinuous | PULSe | TOGGle
*RST Value
CONTinuous
Examples
RES:TRAN:MODE PULS
Query Syntax
[SOURce:]RESistance:TRANsient:MODE?
Returned Parameters
<CRD>
Related Commands
RES:TRAN:ALEV TRAN