Owner manual
60
[SOURce:]VOLTage:TRANsient:BWIDth
Channel Specific
This command specifies the transient pulse width of the input voltage.
Command Syntax
[SOURce:]VOLTage:TRANsient:AWIDth <NRf+>
[SOURce:]VOLTage:TRANsient:BWIDth <NRf+>
Parameters
1 to 65535 us
Unit
S (second)
*RST Value
500uS
Examples
VOLT:TRAN:AWID 0.001 VOLT:TRAN:BWID 0.02
Query Syntax
[SOURce:] VOLTage:TRANsient:AWIDth?
[SOURce:]VOLTage:TRANsient:BWIDth?
Returned Parameters
<NR3>
Related Commands
VOLT
[SOURce:]RESistance
Channel Specific
This command sets the resistance of the electronic load when operating in constant resistance mode.
Command Syntax
[SOURce:]RESistance[:LEVel][:IMMediate] <NRf+>
Parameters
0 to max. rated resistance
Unit
Ω (ohms)
Examples
RES 5 RES:LEV 3.5
Query Syntax
[SOURce:]RESistance[:LEVel][:IMMediate]?
Returned Parameters
<NR3>
Related Commands
RES:RANG
[SOURce:]RESistance: RANGe
Channel Specific
This command sets the resistance range of the electronic load module. This limits the value the
resistance can be set to.
When you program a range value, the electronic load automatically selects the range that corresponds
to the value that you program. If the value falls in a region where ranges overlap, the electronic load
selects the range with the highest resolution.
NOTE: When this command is executed, the IMMediate, TRANsient, TRIGgered, and SLEW resistance