Owner manual

57
Query Syntax
[SOURce:]CURRent:HIGH? [SOURce:]CURRent:LOW?
Returned Parameters
<NR3>
[SOURce:]VOLTage
Channel Specific
This command sets the voltage that the electronic load will regulate when operating in constant
voltage (CV) mode.
Command Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate] <NRf+>
Parameters
See specifications for range
Unit
V (volts)
Examples
VOLT 5
Query Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate]?
Returned Parameters
<NR3>
Related Commands
VOLT:RANG
[SOURce:]VOLTage:RANGe
Channel Specific
This command sets the voltage range of the electronic load module. There is only one voltage range.
Command Syntax
[SOURce:] VOLTage:RANGe <NRf+>
Parameters
see specifications for range
Unit
V (volts)
Examples
VOLT:RANG 15
Query Syntax
[SOURce:]VOLTage:RANGe?
Returned Parameters
<NR3>
Related Commands
VOLT
VOLTage:RANGe:AUTO
Channel Specific
This command sets the voltage auto range state of the electronic load module.
Command Syntax
[SOURce:]VOLTage:RANGe:AUTO<bool>
Parameters
0 | 1 | ON | OFF
*RST Value
1