Owner manual

55
Examples
CURR:PROT 2
Query Syntax
[SOURce:]CURRent:PROTection:LEVel?
Returned Parameters
<NR3>
Related Commands
CURR:PROT:DEL CURR:PROT:STAT
[SOURce:]CURRent:PROTection:DELay
Channel Specific
This command specifies the time that the input current can exceed the protection level before the
input is turned off.
Command Syntax
[SOURce:]CURRent:PROTection[:DELay] <NRf+>
Parameters
0 to 60 seconds
Unit
seconds
*RST Value
0
Examples
CURR:PROT:DEL 5
Query Syntax
[SOURce:]CURRent:PROTection:DELay?
Returned Parameters
<NR1>
Related Commands
CURR:PROT CURR:PROT:STAT
[SOURce:]CURRent:TRANsient:MODE
Channel Specific
This command selects the operating mode of the transient generator as follows in constant current
mode.
CONTinuous The transient generator puts out a continuous pulse stream after receipt of a trigger.
PULSe The transient generator puts out a single pulse upon receipt of a trigger.
TOGGle The transient generator toggles between two levels upon receipt of a trigger.
Command Syntax
[SOURce:]CURRent:TRANsient:MODE <mode>
Parameters
CONTinuous | PULSe | TOGGle
*RST Value
CONTinuous
Examples
CURR:TRAN:MODE TOGG
Query Syntax
[SOURce:]CURRent:TRANsient:MODE?
Returned Parameters
<CRD>
Related Commands
CURR:TRAN:ALEV TRAN
[SOURce:]CURRent:TRANsient:ALEVel
[SOURce:]CURRent:TRANsient:BLEVel