Owner manual
52
This command specifies the timer setting.
Command Syntax
[SOURce:]INPut:TIMer <NRf+>
Parameters
1 to 60000s | MINimum | MAXimum | DEFault
Unit
seconds
*RST Value
10
Examples
INP:TIM:DEL 5
Query Syntax
[SOURce:]INPut:TIMer:DELay? [ MINimum | MAXimum |
DEFault ]
Returned Parameters
<NR3>
Related Commands
INP:TIM
[SOURce:]CURRent
Channel Specific
This command sets the current that the load will regulate when operating in constant current mode.
Command Syntax
[SOURce:]CURRent[:LEVel][:IMMediate] <NRf+>
Parameters
0 through max rated current
Unit
A (amps)
*RST Value
MINimum
Examples
CURR 5 CURR:LEV 0.5
Query Syntax
[SOURce:]CURRent[:LEVel][:IMMediate]?
Returned Parameters
<NR3>
Related Commands
CURR:RANG
[SOURce:]CURRent:RANGe
Channel Specific
This command sets the current range of the electronic load module. There are two current ranges, high
range (model dependent) and low range (model dependent).
When you program a range value, the load automatically selects the range that corresponds to the
value that you program. If the value falls in a region where ranges overlap, the electronic load selects
the range with the highest resolution.
NOTE: When this command is executed, the IMMediate, TRANsient, TRIGgered, and SLEW current
settings are adjusted as follows.
• If existing settings are within new range, no adjustment is made.
• If existing settings are outside new range, the levels are set to the maximum value of the new
range.