Owner manual

47
Channel Specific
When this command is sent and the electronic load is addressed to talk, all the readings stored in the
buffer are sent to the computer.
Query Syntax
TRACe:DATA?
Returned Parameters
{<NR3>, <NR3>…etc,} or {<NR3> <NR3>, <NR3> <NR3>
…etc} (if TRACe:FEED TWO is set)
TRACe:FILTer
Channel Specific
This command is used to select whether the data in buffer is filtered data.
Command Syntax
TRACe:FILTer[:STATe] <BOOL>
Parameters
0 | 1 | ON | OFF
*RST Value
OFF
Examples
TRAC:FILT 1
Query Syntax
TRACe:FILTer[:STATe]?
Returned Parameters
<NR1>
TRACe:DELay
Channel Specific
This command is used to select the delay time for trigger in buffer.
Command Syntax
TRACe:DELay <NRf>
Parameters
0 to 3600s | MINimum | MAXimum | DEFault
Unit
S (second)
*RST Value
0
Examples
TRAC:DEL 1
Query Syntax
TRACe:DELay? [MINimum | MAXimum | DEFault]
Returned Parameters
<NR3>
TRACe:TIMer
Channel Specific
This command is used to select the interval for timer.