Owner manual

38
*RST Value
0.001
Examples
TRIG:TIM 0.25 TRIG:TIM MAX
Query Syntax
TRIGger:TIMer? [ MINimum | MAXimum | DEFault ]
Returned Parameters
<NR3>
Related Commands
ABOR TRIG TRIG:SOUR TRIG:DEL
System Commands
System commands control the system-level functions of the electronic load that are not directly
related to input control or measurement functions.
SYSTem:PRESet
This command returns the instrument to states optimized for front panel operation.
Command Syntax
SYSTem:PRESet
Parameters
None
SYSTem:POSetup
This command is used to select the power-on defaults. With RST selected, the instrument powers up to
the *RST default conditions. With the SAV0 parameter selected, the instrument powers-on to the
setup that is saved in the specified location using the *SAV command.
Command Syntax
SYSTem:POSetup <CRD>
Parameters
RST | SAV0
*RST Value
RST
Examples
SYST:POS RST
Query Syntax
SYSTem:POSetup?
Returned Parameters
<CRD>
Related Commands
*RST *SAV
SYSTem:VERSion?
This query returns the SCPI version number to which the electronic load complies. The value is of the
form YYYY.V, where YYYY is the year and V is the revision number for that year.
Query Syntax
SYSTem:VERSion?
Parameters
None