User Manual

60
To activate this function, configure the settings below:
Configure Config 2 External Ref. Level
Figure 48 - External Reference Voltage Function
13.1.1 AC Couple
AC couple is selected by setting: Configure → Config 1 → Couple → AC. The reference
voltage is proportional to the Instrument voltage.
1 Vdc reference voltage = 30 Vac Instrument
NOTE: Input voltage cannot exceed 0 to 10 V in AC couple mode.
13.1.2 DC Couple
DC couple is selected by setting: Configure → Config 1 → Couple → DC. The reference
voltage is proportional to the source voltage.
±1 Vdc reference voltage = ±42 Vac source
NOTE: Input voltage cannot exceed -10 V to +10 V in DC couple mode.
This function is not supported when Configure → Config 1 → Couple → AC+DC.
13.2 Trigger In
To activate the Trigger function, follow the settings: Configure Config 2 External
Ref. Level.
This instrument has a trigger input on pin 17 of the digital IO interface. Trigger_in is high
at 5 V and low at 0 V. The Instrument will respond to the falling edge of Trigger_in and