29
Output Timer
ON or OFF
Enable/disable the output timer.
Note: This function is available in
display 2 of 3 only.
Timer Setting
(H/M/S)
0 to 99 (hours) : 0 to 59
(minutes): 0 to 59
(seconds)
Set the timing interval of output
timer. When the output is turned
on the timer will count down to 0
and turn the output off.
External Ref.
OFF or LEVEL or AMP
Enable/disable the input of the
external reference signal
OFF: Function disable.
LEVEL: Input DC reference voltage
(0 to ±10 V) via the digital IO
connector to control output
voltage.
AMP: Input reference waveform to
BNC connector (-6 to +6 V) to
control output waveform.
Note: If the input frequency is over
1200 Hz, the amplitude of output
will diminish.
Remote Inhibit
ON or OFF
Remote shutdown function.
Transient
ON or OFF
When output voltage is changed,
the instrument will output a pulse
signal on “Trans” at the digital I/O
terminal.