User Manual
28
Press the arrow key (▲ ▼) to move the cursor to the wanted configuration. Press
ENTER key to change its value, then press ENTER key to confirm.
Figure 27 - Configuration 1
Move the cursor to Range (V) then Press ENTER key. Press 300 V key to put the supply
in the high voltage range, then press Enter key to confirm.
Or use ▲ ▼ keys (or Rotary) to set the value of Range (V) and other configurations. Use
▲ ▼ keys (or Rotary) to choose other configurations.
8.2 Config 2
Configurations listed below can be edited in Configuration 2 page.
Configurations
Range
Description
Waveform Select
A or B
Select to output waveform A or B
Waveform A Type
SINE, SQUA, CSIN (clipping
sine), THD0-29 (harmonic
distortion sine), USER0-4
(user designed waveform)
** SQUA, CSIN (clipping sine), THD
(harmonic distortion sine), USER
(user define waveform) only for
Frequency ≦100Hz
Waveform A Index
(waveform A clip
level (%))
0-29
(0.0 to 100.0%)
When waveform A is THD, set index
When waveform A is CSIN (clipping
sine), set the clip level %
Waveform B Type
SINE, SQUA, CSIN (clipping
sine), THD0-29 (harmonic
distortion sine), USER0-4
(user design waveform)
** SQUA, CSIN (clipping sine), THD
(harmonic distortion sine), USER
(user define waveform) only for
Frequency ≦100Hz
Waveform B Index
(waveform B clip
level (%))
0-29
(0.0 to 100.0%)
When waveform B is THD, set index
When Waveform B is configured as
a CSIN (clipping sine), set the clip
level %