User Manual

28
Press the arrow key ( ) to move the cursor to the wanted configuration. Press
ENTER key to change its value, then press ENTER key to confirm.
Figure 27 - Configuration 1
Move the cursor to Range (V) then Press ENTER key. Press 300 V key to put the supply
in the high voltage range, then press Enter key to confirm.
Or use keys (or Rotary) to set the value of Range (V) and other configurations. Use
keys (or Rotary) to choose other configurations.
8.2 Config 2
Configurations listed below can be edited in Configuration 2 page.
Configurations
Range
Description
Waveform Select
A or B
Select to output waveform A or B
Waveform A Type
SINE, SQUA, CSIN (clipping
sine), THD0-29 (harmonic
distortion sine), USER0-4
(user designed waveform)
** SQUA, CSIN (clipping sine), THD
(harmonic distortion sine), USER
(user define waveform) only for
Frequency 100Hz
Waveform A Index
(waveform A clip
level (%))
0-29
(0.0 to 100.0%)
When waveform A is THD, set index
When waveform A is CSIN (clipping
sine), set the clip level %
Waveform B Type
SINE, SQUA, CSIN (clipping
sine), THD0-29 (harmonic
distortion sine), USER0-4
(user design waveform)
** SQUA, CSIN (clipping sine), THD
(harmonic distortion sine), USER
(user define waveform) only for
Frequency 100Hz
Waveform B Index
(waveform B clip
level (%))
0-29
(0.0 to 100.0%)
When waveform B is THD, set index
When Waveform B is configured as
a CSIN (clipping sine), set the clip
level %