User Manual
92
Programming
Resolution
Voltage
0.1 V
Phase
0.1 degrees
Frequency
Hz (< 100 Hz)
0.1 Hz (>100 Hz)
Accuracy
Voltage
AC
0.2% + 0.2% of F.S.
DC
0.2% + 0.4% of F.S.
Phase
0.15 %
Frequency
± 1 % (45 Hz to 100 Hz)
Measurement
Resolution
Voltage
0.1 V
Current
0.01 A
Power
0.01 W
Frequency
0.01 Hz (<100Hz)
0.1 Hz (>100Hz)
Accuracy
Voltage
AC
0.25 % + 0.25 % F.S.
DC
0.25 % + 0.5 % F.S.
Current
AC
0.25 % + 0.375 % F.S. (rms)
0.4 % + 0.75 % F.S. (Peak)
0.25 % + 0.25 % F.S. (rms)
0.25% + 0.5 % F.S. (Peak)
DC
0.25 % + 3 % F.S.
0.25 % + 2 % F.S.
Power
1 % of F.S. for frequency ≦ 500 Hz
2 % of F.S. for frequency > 500 Hz
Frequency
0.5 %
AC Input
Voltage
190 to 250V
Frequency
47 to 63 Hz
Maximum Power
2500 VA
3800 VA
Maximum Current
13.2 A
20 A
Power Factor
0.98 (typical)
General
Analog
BNC Input
Input Voltage
Range Max.
0 to ± 10 V
Input Impedance
200 kΩ
Bandwidth
1.2 kHz
Storage Memory
10 programs, up to 100 steps total (list mode)
5 memory locations for user-defined waveforms
9 instrument settings
Remote Interface
Analog programming
5
, USB (USBTMC or virtual com),
RS232
5,
GPIB, and Ethernet
Command Response Time
50 ms
Protection
OVP, OCP, OPP, OTP