9830B Series USER MANUAL Model: 9832B, 9833B High Power Programmable AC Power Sources 1
1 Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument. Before applying power to this instrument: Read and understand the safety and operational information in this manual. Apply all the listed safety precautions. Verify that the voltage selector at the line power cord input is set to the correct line voltage.
Do not use this instrument in an electrical environment with a higher category rating than what is specified in this manual for this instrument. You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the instrument's category rating to maintain the instrument's category rating. Failure to do so will lower the category rating of the measuring system. Electrical Power This instrument is intended to be powered from a CATEGORY II mains power environment.
Do not operate the instrument in the presence of flammable gases or vapors, fumes, or finelydivided particulates. The instrument is designed to be used in office-type indoor environments. Do not operate the instrument In the presence of noxious, corrosive, or flammable fumes, gases, vapors, chemicals, or finely-divided particulates. In relative humidity conditions outside the instrument's specifications.
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual. Not for critical applications This instrument is not authorized for use in contact with the human body or for use as a component in a life-support device or system. Do not touch live circuits Instrument covers must not be removed by operating personnel.
instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a fire. Failure to use the specified fuses will void the warranty. Servicing Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to B&K Precision for service and repair to ensure that safety and performance features are maintained. For continued safe use of the instrument Do not place heavy objects on the instrument.
Compliance Statements Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with separate collection systems) This product is subject to Directive 2002/96/EC of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste.
Safety Symbols Refer to the user manual for warning information to avoid hazard or personal injury and prevent damage to instrument. Electric Shock hazard Hot surface On (Power). Press the “I” at the top of the power switch to turn the instrument ON. Off (Power). Press the “O” at the bottom of the power switch to turn the instrument off.
CAUTION indicates a hazardous situation which, if not avoided, will result in minor or moderate injury WARNING indicates a hazardous situation which, if not avoided, could result in death or serious injury DANGER indicates a hazardous situation which, if not avoided, will result in death or serious injury. NOTICE is used to address practices not related to physical injury.
2 Notations TEXT – Denotes a softkey. TEXT – Denotes a front panel key.
Table of Contents 1 Safety Summary ............................................................................................................... i Compliance Statements............................................................................................................... vi Safety Symbol ............................................................................................................................. vii 2 Notations ....................................................................
8.1 8.2 8.3 9 Config 1 ............................................................................................................................ 27 Config 2 ............................................................................................................................ 28 Limits................................................................................................................................ 30 System Settings ....................................................................
.3 15 VDC ............................................................................................................................. 61 13.4 Output Status Detection .................................................................................................. 62 13.4.1 /SYNC ................................................................................................................. 62 13.4.2 /Fault_out ............................................................................................
3 General Information The B&K Precision 9830B series are low distortion single phase AC power sources delivering a maximum of 3000 VA, 300 Vrms, 30 Arms /97.5 Apk. The Output frequency is adjustable from 45 Hz to 1200 Hz. All models are capable of outputting AC, DC or AC+DC. Predefined waveforms include sine, square, clipped sine and THD waveforms. Standard remote interface include USBVCP and USBTMC-compliant. RS232, LAN, and GPIB interfaces are available to provide flexibility for remote operation.
Note: Check that you have the most current User manual. This is available for download at www.bkprecision.com 3.
3.3 Installation The following diagram shows how to install the optional rack mount kit.
3.
Key Definition Frequency key, press to set the output frequency Voltage key, press to set the output voltage Press to lock the keyboard, (Keyboard is locked when LED is illuminated). Output On/Off key, press to enable/disable voltage output, (Output is On when LED is illuminated). to Number key 1 to 9 for direct numeric entry. Set the number entered to a positive or negative value.
3.
4 Display Overview The power on screen shows the model number and self-test status. The Interface, and Power module should indicate OK. The (Real Time Clock), RTC will show the battery voltage. When the self-test completes, the instrument will advance to the Settings and Measure screen. This process takes approximately 7 seconds. If an error does occur, the instrument must be returned for service. Figure 5 - Power on screen 4.
Figure 6 - Display Description The instrument provides 3 display screen options for accessing commonly used measurements and settings. Each of these screens can be selected by pressing the Display x of x softkey repeatedly.
Figure 8 - Display 2 of 3 Display 3 of 3 This displays shows a graphical representation of the settings, output measurements and waveforms. Figure 9 - Display 3 of 3 Press the ◄ ► keys to select and view V (voltage waveform) or A (current waveform). Then press the ▲ ▼ key to change the scale until the complete waveform is visible.
5 Getting Started Before connecting and powering up the instrument, please review and go through the instructions in this chapter. 5.
current required to operate the instrument. 2. Using the power cable provided, identify the end that plugs into the instrument. Before disassembling the black hooded connector, make a note of how the unit is assembled from the factory so it can be correctly reassembled. 3. Remove the 4 screws from the bottom of the hooded connector as shown and place the upper and the upper and lower hood pieces and screws to the side. Remove 4 screws The black, white, and green wires should now be clearly visible.
up and reinstall the 4 screws into the hood. The green connector should be locked in place. Caution Label Latch Fingers Key section 4. Using the (2) screws provided with the cable, secure the assembled connector to the back of the instrument as shown in Figure 10 – Securing. Note: Use only the screws provided. Key Section Figure 10 - Securing the power cable The green (trace yellow) wire connects to electrical ground (required).
instrument. The AC voltage range must meet per the acceptable specifications. 4. Connect Power 5. Connect specified AC power cord and verify the hood is in place and correctly secured to rear panel. 6. Press the power switch to the “|” ON position to turn ON the instrument. It will take a few seconds before the fans power on and the self-test begins. 5.2.2 Warm-up Time The 9830 series is fully operable upon switching the power on.
as short as possible. The local sense is the default configuration. When remote sense is selected, both the sense end L and output end L are connected to the load end L, whereas both sense end N and output end N are connected to the load end N.
6 Menu Tree 15
7 Front Panel Operation 7.1 Configure Voltage and Frequency Output 7.1.1 Setting voltage Press VOLT or the ▲ ▼ (arrow keys) to move the cursor to Volts (rms). Then press the ENTER key. There are three ways to set the value of the output voltage: 1. Press the number keys 1, 1, 0, . , 0 and Enter to set Volts(rms) = 110.0 V. 2. Use the ◄ ► keys to select the hundreds digit then press the arrow keys ▲ or ▼ repeatedly until the value reaches 110.0 and press the Enter or OK key to confirm. 3.
Figure 12 - Setting Frequency 7.1.3 Setting Voltage Press the VOLT key, and a small window will pop-up. Pressing the VOLT key again will switch between VAC and VDC (when the coupling mode is “AC+DC”). Press the number keys to set the value, and press the OK or the Enter key to confirm the change. Optionally, press the ESC key to cancel the setting change. Vac Vdc Figure 13 - Setting Voltage. Vac left, Vdc right.
7.2 Program Settings Press the Program softkey in the Setting and Measurement page to enter Program mode. There are three modes for the user to choose from to simulate Power Line Disturbances. They are: Step Output settings step-up or step-down based on user criteria. List Change the output sequentially by individual settings included in a list. Pulse Trigger pulse signals periodically. 7.2.1 Step Mode Step Mode has 10 parameters that can be set by the user.
Step Mode Operation Steps The following example for Step Mode waveform shows 4 steps starting at 40 volts and increasing 20 volts at each step. The instrument will dwell for 80 ms at each step. Figure 14 - Step Mode Example Output Parameters Volts(rms) dVac Volts(DC) dVdc Freq. dF Set Time Count Sync Source Phase(deg) Settings 40 +20 +0 0 50 0.0 80 4 Phase 0.0 Table 3- Step Mode Example Parameters Press the Step softkey.
1. Use the arrow keys (▲ ▼ ◄ ►) to move the cursor to the wanted parameter of the STEP setting page. 2. Press the ENTER key to adjust the parameter. Then press the ENTER key again to change its value. 3. When finished editing all the parameters, press On/Off key to activate the Step mode. Using an oscilloscope you should observe a wavfore as shown in Figure 16 - Oscilloscope view of Step waveform example.
Parameters List No. Step No. Volts (rms) Start Volts (rms) End Volts (DC) Start Volts (DC) End Freq. (Hz) Start Freq. (Hz) End Time (ms) Range 0 to 9 0 to 99 0 to 300 0 to 300 -424.0 to 424.0 -424.0 to 424.0 43 to 1200.0 43 to 1200.
List operation pages Press the List softkey, refer to List Parameter Table for settings details. Figure 18 - List Mode first page Refer to Table 6 – List Mode First Page. The first page parameters are common to all the steps added under the second page parameters. This first page only needs to be configured once for List 0. Use the arrow keys (▲ ▼ ◄ ►) to move the cursor to the wanted parameter of the List page. Then press the ENTER Key to change its value.
yellow boxes when no entries are stored in any List. Figure 19 - List Mode second page Refer to Error! Reference source not found.. Use the arrow key (▲ ▼) to move the cursor to the wanted parameter of List Step Configuration page. Press the ENTER key to change its value, then press the ENTER key again to confirm the change. Parameters Volts (rms) Start Volts (rms) End Volts (DC) Start Volts (DC) End Freq. (Hz) Start Freq.
. Use the arrow keys (▲ ▼) to move the cursor to the wanted parameter of List Step Configuration page. Press the ENTER key to change its value, then press the ENTER key again to confirm the change. Parameters Volts (rms) Start Volts (rms) End Volts (DC) Start Volts (DC) End Freq. (Hz) Start Freq.
changes in to memory and add a new step in List No. 0. The page with Empty Step field in yellow will change to 97 and Current Step to 3. All of the steps 0, 1 and 2 have been entered. Press the ESC key to exit. Make sure Set Repeat is = to 1 and press On/Off key to activate the List mode. Figure 23 - Oscilloscope view of List mode waveform example 7.2.3 Pulse Mode The following example will describe how to setup the Pulse Mode. Parameters Volts (rms) Volts (DC) Freq. Duty Range 0 to 300.0 Vrms -424.
Pulse Mode Operation Steps 1. Press Pulse softkey. 2. Use the arrow key (▲ ▼) to move the cursor to the wanted parameter. 3. Press the ENTER key to change its value then press the ENTER key again to confirm. Figure 24 - Pulse mode page 4. When finishing editing the parameters press the On/Off key to active the Pulse mode.
8 Configure Menu In Setting and Measure page, press Configure key to open the Configuration page. The Configuration page includes three softkey options: 1. Config 1 (default) 2. Config 2 3. Limits Each of these pages will be explained below with more detail. Figure 26 - Default Configuration 1 Menu 8.1 Config 1 Configurations listed below can be edited in this Configuration 1 page.
Press the arrow key (▲ ▼) to move the cursor to the wanted configuration. Press ENTER key to change its value, then press ENTER key to confirm. Figure 27 - Configuration 1 Move the cursor to Range (V) then Press ENTER key. Press 300 V key to put the supply in the high voltage range, then press Enter key to confirm. Or use ▲ ▼ keys (or Rotary) to set the value of Range (V) and other configurations. Use ▲ ▼ keys (or Rotary) to choose other configurations. 8.
Output Timer ON or OFF Timer Setting (H/M/S) 0 to 99 (hours) : 0 to 59 (minutes): 0 to 59 (seconds) External Ref. OFF or LEVEL or AMP Remote Inhibit Transient ON or OFF ON or OFF 29 Enable/disable the output timer. Note: This function is available in display 2 of 3 only. Set the timing interval of output timer. When the output is turned on the timer will count down to 0 and turn the output off. Enable/disable the input of the external reference signal OFF: Function disable.
Press the arrow key (▲ ▼) to move the cursor to the wanted configuration. Press the ENTER key to change its value then press the ENTER again key to confirm. See the example below: Move the cursor to Waveform Select, then press OK or Enter. Press the A softkey then press OK or Enter to confirm. Use ▲ ▼ keys (or Rotary knob) to select the other configurations. Figure 28 - Configuration 2 page 8.3 Limits The Configuration Limits page is used to set user-defined voltage, current and power limits.
Press the arrow key (▲ ▼) to move the cursor to the wanted configuration. Press ENTER key to change its value, then press ENTER key to confirm. Figure 29 - Configure Limits Move the cursor to Volts (rms). Use ▲ ▼ keys (or Rotary knob) to set the value of Volts (rms) limit and other configurations. Then press OK or Enter key to confirm this setting. Use ▲ ▼ keys (or Rotary knob) to select other configurations.
9 System Settings 9.1 System Setup The System Setup page is used to set Date, Time, Brightness and Beep. Figure 30 - System Setup Press the arrow keys (▲ ▼) to move the cursor to the wanted configuration. Press the ENTER key to change its value and then press the ENTER key again to confirm the changes.
9.2 Communication Setup The Communication Setup page is used to select and configure the communications ports. Figure 31 - Communication Setup Press the arrow keys (▲ ▼) to move the cursor to the wanted configuration. Press the ENTER key to change its value and then press the ENTER key again to confirm the changes. Configurations Comm.
9.3 System Error The System Clear Error page is used to view and clear the error log. Use + - key or rotary knob to see other error messages. Figure 32 System Error Log To clear the error log Press the arrow key (▲ ▼) to move the cursor to Clear Error? (Yes/No) then press the ENTER key to change its value. Select YES, then press ENTER key again to clear error log. The log errors cannot be recovered once it has been cleared.
Table 15 - Error Code Definitions 9.4 System Next The System Next page is used to access to system defaults and calibration functions. Figure 33 - System Next Page 9.5 Recall Default Press System softkey, then press Next softkey to go to the next function page. At the System Next screen press Recall Default softkey. Press an arrow key (▲ ▼) to move the cursor to Reset all config? Press ENTER key, then press the Yes softkey to reset all configurations.
(User) Freq(Hz) (User) Sync Source (User) Phase(deg) (Inrush) Measurement Time(ms) (Inrush) Delay Time(ms) Waveform Select Waveform A Type Wave A Index Waveform B Type Wave B Index Wave B Output Timer Timer Setting External Ref.
10 Save The Instrument can save data and screen shots into the USB disk (only FAT format is supported). Figure 35 - Save Menu 10.1 Save Config Press Save Config softkey to store all configurations and settings into the internal memory (CFGFile01.cfg to CFGFile09.cfg) or into a USB disk (CFGFile10.cfg to CFGFile99.cfg). Use the number keys to enter the file name. Then press OK or ENTER key to confirm or press ESC key to cancel.
10.2 Save Screen Plug in a USB disk, then press the Save Screen softkey to take a screen shot and store the image to a USB disk (SCRFile000.bmp to SCRFile999.bmp). Use number keys to enter the file name. Then press OK or ENTER key to confirm or press ESC key to cancel. Figure 37 - Save Screen 10.
Press Recall Config softkey to recall configurations and settings from internal memory (CFGFile01.cfg to CFG09File.cfg) or from a USB disk (CFGFile10.cfg to CFGFile99.cfg). Use the number keys to enter the file name. Then press OK or ENTER key to confirm, or press ESC key to cancel.
11 3-Phase Operation Three-Phase or Multi-Phase operation requires the three 9832B or 9833B AC sources and the 3Phase kit (model TL983P-KIT) which includes three phase sync modules and two straight through (pin to pin) RJ45 communication cables for 3-Phase operation. Replacement modules (model TL983P) are sold separately. Additional phases require additional modules and RJ45 communication cables.
RJ45 cables Additional phases require additional AC sources and TL983P modules. 2) Front panel configuration After verifying the TL983P adapters are properly wired on the rear panel, the AC sources must be configured from the front panel. 3-phase operation is enabled by setting the phase control field on each AC source. The options for phase control are Master, Slave, or Disable. The master must be set first. Any one of the three connected AC sources can be set to master.
Similarly, after setting the master set ‘Phase Ctrl’ of the remaining two AC sources to ‘Slave’. The slaves will automatically connect to the master. This completes the set-up process for three phase operation. 3) 3-Phase parameters and output 3-Phase parameters include: Voltage (rms), Frequency (Hz), and Phase (deg). The reference phase of the master is always fixed at 0 degrees. The phase angle of slave sources can be adjusted from 0 to 359.7 degrees.
11.3 3-Phase Output Wiring The 9830B Series supports a Y-configuration in 3-phase mode. Connect the N(-) terminals on the rear panel together as shown in Figure 11.
11.4 Phase Control Software The 9830B Series Phase control software can be downloaded from bkprecision.com by visiting the 9830B product page and clicking the ‘Docs & Software’ tab. Phase control software allows you to easily control 9832B/9833B models in three-phase or multi-phase modes with data logging and live output monitoring capabilities. 11.4.
Three-Phase Application Mode Control Panel Multi-phase Application Mode This mode is intended for applications requiring control over more than three phases or when the user wants independent control of each AC source Voltage (Vrms) and Phase (degrees) parameters. Up to 31 instruments can be connected in multi-phase application mode.
11.4.2 Setup and Configuration Before attempting to connect to a computer, the 9832B/9833B AC Power Sources must be properly configured from the front panel and connected using phase sync adapter modules. See the 3-Phase Operation section above for instructions. On the front panel, set each AC source to ‘USBVCP’ in the communication settings. Figure 40 - Communication Setup Press the arrow keys (▲ ▼) to move the cursor to the wanted configuration.
1. Open the phase control software and click the icon to open the windows device manager. Three connected AC Sources should be displayed under ‘Ports (COM & LPT)’ in the device manager as shown: If they are missing or shown with an error power off and disconnect the AC Sources from the computer. Power back on and connect AC Sources to the computer one at a time. 2. Click ‘Refresh’ and the software will automatically search for connected AC sources and assign one master and two slaves.
Click ‘Set’ to send the parameters and prepare the 3-Phase system for output. Output Click ‘Output On’ to enable the output of the 3-Phase system. Fields will be highlighted in green when the output is enabled. The Output Monitor panel displays live Voltage (Vrms), Current (Arms), Power (VA), and Power (W). Click ‘Output Off’ to disable the output of the system.
Check the box to enable data logging and set the desired sampling interval. The sampling interval is displayed in seconds (s). Data logging will start when the output is enabled. Output Timer The output timer is used to disable the output after a certain period of time. The timer can be set from 1 second to over 277 hours. To enable output timer, click ‘Setting’ and the following menu will appear: Check the box under ‘Timer setting’ to enable the timer.
1. Click the icon to open the windows device manager. Three connected AC Sources should be displayed under ‘Ports (COM & LPT)’ as shown: If they are missing or shown with an error power off and disconnect the AC Sources from the computer. Power back on and connect AC Sources to the computer one at a time. 2. Click ‘Scan’ and the software will automatically connect to and assign AC sources for multi-phase operation. There will be one master with the remaining connected AC sources set to slave.
Click ‘Output On’ to enable the output of the multi-phase system. Fields will be highlighted in green when the output is enabled. The Output Monitor panel displays live Voltage (Vrms), Current (Arms), Power (VA), and Power (W). Click ‘Output Off’ to disable the output of the system. Data Logging This software is capable of logging time, voltage (Vrms), current (Arms), power (VA), and power (W) at user-defined sampling intervals for each AC Source. Data is exported as a .CSV file.
Check the box to enable data logging and set the desired sampling interval. The sampling interval is displayed in seconds (s). Data logging will start when the output is enabled. Output Timer The output timer is used to disable the output after a certain period of time. The timer can be set from 1 second to over 277 hours. To enable output timer, click ‘Setting’ and the following menu will appear: Check the box under ‘Timer setting’ to enable the timer.
12 Remote Interface Operation The Instrument comes with RS232, USB (USBTMC, and USBVCP), GPIB, LAN and analog interfaces. Users can program the Instrument using the SCPI (Standard Commands for Programmable Instruments) commands through any of the remote interfaces. Only one interface at a time can be enabled and used to control the Instrument. 12.1 Interface Connection 12.1.1 RS232 The RS232 interface can be found on pin 23 (RX) and 11 (TX) of the DB25 port.
12.1.3 USBTMC The USB port is USBTMC-compliant and can be used for remote communication and control. There are no additional settings in the menu system for USB configuration. The only requirement is that the USBTMC driver be installed. It is included when installing VISA software on the computer. (We recommend using NI-VISA, which can be downloaded at http://www.ni.com/visa/). 12.1.4 GPIB Each device is assigned a GPIB address from 1-30.
Figure 41 - Web Login Page A password is required to login and access the menu items on the page. The default admin password is 123456. The web server menu items are described below: HOME This page provides general information about the instrument, Manufacturer, Model Number, Serial Number, Firmware version, Interface, USBTMC setting, MAC address, and IP Address.
LAN Config This page provides settings and status of LAN, including IP address, IP subnet, Gateway, DNS server, Hostname, Domain, mDNS host name, TCP/IP VXI-11 instrument, and TCP/IP Raw Socket. Figure 43 - LAN Config Config The Config page provides different settings output voltage range, waveform A/B, output type, inrush current. Limitations can also be set for R.M.S. voltage, +/- Peak voltage, output R.M.S. current and output power.
Control The Control page provides the general control of the instrument such as output on/off as well as the AC/DC voltage and frequency settings. The command line for SCPI commands can also be accessed here. Figure 45 - Source Control Log Out The Log Out will exit the web page and go back to login screen. The instrument can be connected via LAN (Ethernet) or Telnet client with a socket port of 5024 Socket connection. Socket connection is available for communication via LAN (Ethernet) interface.
13 Digital I/O The digital I/O interface is used to control or monitor the Instrument. Refer to the figure below describing the digital I/O 25-pin. Figure 46 - Digital IO Pinout Pin 1 Name Ext_V I/O IN Definition External reference voltage input pin. (reference ground: AGND) In DC mode, enter -10V to +10V to control the DC output. In AC mode, enter 0 to +10V to control the AC output. Not supported in AC+DC mode. +15VDC output.
13 Event_SW+ IN 14 15 16 AGND NONE /Fault_out PWR 17 /Trigger_in IN 18 19 DGND /Remote_inhibit PWR IN 20 21 DGND /Transient PWR OUT 22 23 24 25 DGND Rx NONE Event_SW- PWR IN OUT IN Built-in switch + (control the switch on or switch off via the SCPI commands) Analog ground When protection status is active, the voltage level of this pin will turn from high to low.
To activate this function, configure the settings below: Configure → Config 2 → External Ref. → Level Figure 48 - External Reference Voltage Function 13.1.1 AC Couple AC couple is selected by setting: Configure → Config 1 → Couple → AC. The reference voltage is proportional to the Instrument voltage. 1 Vdc reference voltage = 30 Vac Instrument NOTE: Input voltage cannot exceed 0 to 10 V in AC couple mode. 13.1.2 DC Couple DC couple is selected by setting: Configure → Config 1 → Couple → DC.
Ext_OnOff (external output control) according to the following control table and timing diagram. /Trigger_in Falling Edge Falling Edge HIGH HIGH Falling Edge HIGH /Ext_OnOff LOW HIGH LOW HIGH LOW LOW Result Instrument output STOP Instrument output ON No change No change Instrument output STOP Instrument output ON Table 19 - Control Table Figure 49 - Trigger Timing Diagram 13.3 15 VDC This instrument supplies 15 VDC up to 100 mA between pin 2 (15VDC) and pin 14 (AGND).
13.4 Output Status Detection The digital IO interface has system status logic outputs that are described in this section. All these outputs are in reference to DGND, pins 8, 10, 18, 20, and 22. Figure 51 - Output status pin out 13.4.1 /SYNC When the output sine wave of the Instrument is about to pass through 0 degrees, pin 9 (/SYNC) will send out a 250 uS 12 V to 0 V pulse as shown in the below timing diagram.
13.4.2 /Fault_out Pin 16 (/Fault_out) indicates when a fault occurs or the Instrument is in protection status. During normal operation, the voltage level of this pin stays high (5 V). The voltage level will be low (0 V) if any of the circumstances below occurs: 4. Interface CAN Error 5. Module CAN Error 6. SW Over Current Prot 7. SW Over Voltage Prot 8. SW Over Power Limit 9. SW Over Max VA Limit 10. Remote Inhibit ON 13.4.
13.5 Remote_Inhibit The Instrument offers a remote inhibit used to turn the output off. The voltage level of pin 19 (Remote_inhibit) (reference ground is DGND) has to be set high (5 V) first, or the Instrument will enter into protection mode. If the voltage level drops to 0 V, the Instrument will stop outputting, and will show ‘Error Remote Inhibit ON’.
The RS232 interface does not support flow control. The programmer should be aware of this limitation and pay attention to the Instrument command process time. If the remote commands are sent too fast the internal buffer may overrun and cause communication errors. Therefore, it is mandatory to add a delay between commands so that the Instrument has sufficient time to process. 13.7 Event_SW The Event_SW output can be used to control external devices.
13.8 Analog input (BNC) The analog input(BNC) is used to control the output voltage of the instrument with an external signal. In this example, an Arbitrary waveform generator is connected.
Press Configure softkey in the Setting and Measurement screen. Configure as follows: Config 1(default) → Config 2 → Limits Press Config 2 softkey to enter the Configuration 2 setting page. Use the arrow key (▲ ▼) to move the cursor to the External Ref., then press the ENTER key to change value. Select the Amp softkey and press ENTER to save the setting. Figure 59 - Configuration 2 Amp mode Press ESC to go back to the Setting and Measurement screen.
14 Built-in Harmonic Wave THD 00 Harmonic 2 5 weight(%) 2.07 9.8 7 8 15.8 2.16 THD 01 Harmonic 3 7 19 weight(%)) 1.5 1.5 2 Harmonic 3 5 7 23 31 weight(%) 2 14 2 1.
THD 03 Harmonic 3 5 7 23 25 31 33 weight(%) 2.5 1.9 2.5 1.9 1.1 1.5 1.1 Harmonic 3 5 7 9 11 weight(%) 1.1 2.8 1.4 2.3 1.
THD 05 Harmonic 3 5 weight(%) 1.65 4.2 7 15 19 3.45 1.05 3 THD 06 Harmonic 3 5 7 9 11 15 21 weight(%) 2.2 5.6 2.8 4.6 3 1.
THD 07 Harmonic 3 5 7 11 15 17 weight(%) 4.9 1.6 2.7 1.4 2 1.1 5 7 11 13 15 THD 08 Harmonic 3 weight(%) 7.35 2.4 4.05 2.1 1.05 3 17 19 21 23 1.65 1.05 1.05 1.2 71 25 1.
THD 09 Harmonic 3 5 7 9 11 13 15 17 19 21 23 25 weight(%) 9.8 3.2 5.4 1.2 2.8 1.4 4 2.2 1.4 1.4 1.6 1.4 THD 10 Harmonic 3 weight(%) 17.
THD 11 Harmonic 3 weight(%) 21.3 THD 12 Harmonic 3 weight(%) 24.
THD 13 Harmonic 2 5 7 8 weight(%) 2.3 9.8 15.8 2.5 5 7 8 7.9 1.25 THD 14 Harmonic 2 weight(%) 1.15 4.
THD 15 Harmonic 5 7 weight(%) 2.45 3.95 THD 16 Harmonic 3 5 7 weight(%) 11 4.05 2 9 1.
THD 17 Harmonic 3 5 9 weight(%) 7.17 3.42 0.8 THD 18 Harmonic 3 5 9 weight(%) 8.11 3.
THD 19 Harmonic 3 5 9 weight(%) 9.38 3.44 1.15 THD 20 Harmonic 3 5 7 9 11 weight(%) 2 1.8 1.6 1.23 0.
THD 21 Harmonic 3 5 weight(%) 3 7 9 11 13 2.75 2.4 2 1.4 0.8 5 9 11 13 2 1.25 THD 22 Harmonic 3 weight(%) 4.15 3.8 7 3.24 2.
THD 23 Harmonic 3 5 7 9 11 13 15 21 23 25 27 weight(%) 5.63 5.13 4.42 3.56 2.63 1.68 0.79 1.04 1.27 1.32 1.2 29 0.95 THD 24 Harmonic 3 5 7 9 11 13 15 21 23 25 27 29 weight(%) 7.28 6.63 5.71 4.61 3.42 2.19 1.04 1.32 1.63 1.69 1.54 1.
THD 25 Harmonic 5 7 11 13 19 23 25 35 37 weight(%) 3.54 2.68 8.87 7.86 1.04 4.11 4.13 2.61 2.82 THD 26 Harmonic 21 23 25 weight(%) 1.38 5.39 2.
K THD 27 Harmonic 3 weight(%) 33 20 14 11 8.5 7.2 6 5 7 9 11 13 15 17 19 21 5 5 23 25 4.5 4 27 29 31 33 35 37 39 3.5 3 2.5 2 2 THD 28 Harmonic 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 weight(%) 33 20 14 11 8.5 7.2 6 5 5 4.
THD 29 Harmonic 3 5 weight(%) 33.3 20 7 9 11 13 15 13.8 10.8 8.5 7.2 5.
15 Calibration B+K Precision recommends a calibration interval of one year for this instrument. To perform the calibration, the following equipment is required: 5 1/2 digital multimeter (DMM); B+K Precision 5492B or equivalent DC source: BK Precision 9110 or equivalent Resistive load: 3 ohm with rated power over 5,600 watts WARNING Calibration must only be performed by authorized technicians or calibration personnel only.
Calibration is performed for both the 150V range and the 300V range. The following steps repeat for each range: 15.1.1 300V range calibration 1. Press Configure in the Setting and Measurement screen. Select 300V for the Range (V). Return to Setting and Measurement screen and select System → Next → Calibration and enter the password 13579. 2. Select Vmeas (AC voltage) on the calibration menu.
The calibration date and time will be shown at the bottom of the screen. Next, calibrate Freq2, Freq3 and Freq4 successively to complete the Range(V) 300V calibration. 15.1.2 150V range calibration Press ESC key to return to Setting and Measurement screen and press Configure. Select 150V for Range (V). Return to the Setting and Measurement screen, and repeat Steps 2 ~ 3 to calibrate Freq1 through Freq4. Note that the HIGH calibration voltage of 260VAC becomes 140VAC. 15.1.
“OK” after the Low Point(V). 2. Press Cal. softkey to start calibration. The instrument will output ML Point voltage. Use the number keys to enter the voltage shown on the DMM. Use the number keys to enter the voltage shown on the DMM. Press ENTER, and the instrument will show “OK” after the ML Point(V). 3. Press Cal. softkey to start calibration. The instrument will output MH Point voltage. Use the number keys to enter the voltage shown on the DMM. Use the number keys to enter the voltage shown on the DMM.
1. Connect the DMM to both ends of the shunt. Set the DMM to measure AC voltage. 2. Press the System → Next → Calibration → softkeys. Enter the password 13579. 3. Then select Imeas (AC current) on the calibration menu. Freq1, Freq2, Freq3 and Freq4 represent output frequency of 100Hz, 400Hz, 800Hz and 1200Hz respectively. 5. Press Freq1 key then press the Cal. to start calibration, and the instrument will output the Low Point current.
15.1.6 External Voltage Calibration Connect a DC power supply to pin 1 and pin 14 of Digital IO. The DC power must be able to supply 10V. CAUTION: The DC source for external voltage control and the DC source for external voltage calibration should be identical. And the external voltage calibration should be completed before using external voltage control function. Press System → Next → External Calibration softkeys in Setting and Measurement screen.
4. Input 9V from the external DC power supply. The ADC value should be 14900±300, then press the Next point softkey to calibrate the second point. 5. Supply 0V from the external DC power supply. The AD value should be 7900±300, then press the Next point softkey to calibrate the third point. 6. Input -9Vdc from the external DC power supply. The AD value should be 1000±300, then press the Next point softkey.
1. The instrument will show “Cal End”, then press ESC to complete the external voltage calibration.
16 Specifications Note: All specifications apply to the unit after a temperature stabilization time of 15 minutes over an ambient temperature range of 23 °C ± 5 °C. Specifications are subject to change without notice. Model AC Output Output Phase Max. Power Voltage Range1 (rms) Current (rms) Current (peak) Low High Low High Low High Frequency Range Single phase 3-Phase Phase Range Total 45 to 400 Hz Harmonic > 400 to 1 KHz 2 Distortion > 1 k to 1.
Programming Resolution Voltage Phase Frequency Voltage Accuracy Phase Frequency 0.1 V 0.1 degrees Hz (< 100 Hz) 0.1 Hz (>100 Hz) AC 0.2% + 0.2% of F.S. DC 0.2% + 0.4% of F.S. 0.15 % ± 1 % (45 Hz to 100 Hz) Measurement Resolution Voltage Current Power Frequency Voltage Accuracy Current Power Frequency AC Input Voltage Frequency Maximum Power Maximum Current Power Factor General 0.1 V 0.01 A 0.01 W 0.01 Hz (<100Hz) 0.1 Hz (>100Hz) AC 0.25 % + 0.25 % F.S. DC 0.25 % + 0.5 % F.S. 0.25 % + 0.375 % F.S.
Operating Temperature Storage Temperature Environmental Conditions Dimensions ( W×H×D ) Weight Included Accessories Optional Accessories 32 °F to 104 °F (0 °C to 40 °C) -40 °F to 185 °F (-40 °C to 85 °C) 80% Relative Humidity up to 35 °C, non-condensing 16.5” x 5.2” x 22” (420 x 132 x 560 mm) 52.9 lbs.
17 LIMITED THREE-YEAR WARRANTY B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of three years from date of purchase. B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt.
22820 Savi Ranch Parkway Yorba Linda, CA 92887 www.bkprecision.com © 2020 B&K Precision Corp.