User Manual
176
Figure 32- Current Calibration with Shunt
Follow the steps below to proceed with current calibration:
1. From the calibration menu, press to enter CURRENT
CALIBRATION. The following screen will be shown.
2. Select the channel (for dual channel models), then press .
The calibration process will automatically start.
CURRENT CALIBRATION CH: 1
IL Set = X . XXXXA Read = XXX
IML Set = X . XXXXA Read = XXX
Enter
1
6 ½ Digit DMM