User Manual
172
Remote Sense
Compensation
1V Max
Rising Time at Full Load
≤ 8ms (9171B,9173B,9181B,9182B)
≤ 10ms (9172B,9174B,9183B)
≤ 30ms (9184B)
≤ 40ms (9185B)
Rising Time at No Load
≤ 8ms (9171B,9173B,9181B,9182B)
≤ 10ms (9172B,9174B,9183B)
≤ 30ms (9184B)
≤ 40ms (9185B)
Falling Time at Full Load
≤ 8ms (9171B,9173B,9181B,9182B)
≤ 10ms (9172B,9174B,9183B)
≤ 30ms (9184B)
≤ 40ms (9185B)
Falling Time at No Load
≤ 250ms
Standard Interface
USB
Optional Interface Cards
5
LAN/GPIB Card, Digital I/O and Analog
Control Cards (1CH and 2CH), RS485
Card, RS232 Card
Remote Command
Processing Time
≤ 10ms (typical)
General
AC Input
115/230 VAC± 10%, 47Hz - 63Hz
Operating Temperature
0°C- 40°C
Storage Temperature
-10°C- 70°C
1
With sense terminal connected.
2
Maximum time required for the output voltage to change from 1% to 99% or
vice versa following the receipt of VOLTage or VSET command via direct GPIB
or USB interface.
3
Models 9171B – 9183B: Less than 50 μs for output to recover to within 15 mV
following a change in output current from full load to half load or vice versa.
Model 9184B: Less than 100 μs for output to recover to within 50 mV following
a change in output current from full load to half load or vice versa.
Model9185B: Less than 100 μs for output to recover to within 120 mV following