User Manual

146
0
7
6
5
4
3
2
1
0
Channel 1 OVP On/Off status
Channel 2 OVP On/Off status
Channel 1 OCP On/Off status
Channel 2 OCP On/Off status
Channel 1 output On/Off status
Channel 2 output On/Off status
LCD backlight On/Off Status
Output mode Single/Multi status
1 is ON, 0 is OFF
1 is ON, 0 is OFF
1 is ON, 0 is OFF
1 is ON, 0 is OFF
1 is ON, 0 is OFF
1 is ON, 0 is OFF
0 Always ON, 1
Timed
0 – MULTI, 1 – SINGLE
Example 2:
Return string: 00008C
Byte 2: 0
Byte 1: 0
Byte 0: 8C
8C = 10001100 (bits 7 – bit 0)
Representation:
- Channel 1 OVP status is ON
- Channel 1 output is ON
- Channel 2 output is ON
- LCD backlight is ALWAYS ON
- Output mode is MULTI
- No protection flag status (instrument is operating normally)
TRACK
Note:
This command is for dual channel models only.
Description:
Configures/Queries the tracking mode state.
Syntax:
TRACK<SP><state>
TRACK?
<state> - ON or 1 to turn ON
- OFF or 0 to turn OFF