Specifications 9171B 9172B 9173B 9174B 9181B 9182B 9183B 9184B 9185B
Output Rating
Low Range
0-10 V,
0-10 A
0-35 V,
0-3 A
0-10 V,
0-10 A
0-35 V,
0-3 A
0-18 V,
0-8 A
0-10 V,
0-20 A
0-35 V,
0-6 A
0-100 V,
0-2 A
0-400 V,
0-0.5 A
High Range
0-20 V,
0-5 A
0-70 V,
0-1.5 A
0-20 V,
0-5 A
0-70 V,
0-1.5 A
0-36 V,
0-4 A
0-20 V,
0-10 A
0-70 V,
0-3 A
0-200 V,
0-1 A
0-600 V,
0-0.35 A
Number of Channels 1 1 2 2 1 1 1 1 1
Max Output Power 100 W 105 W 200 W 210 W 144 W 200 W 210 W 200 W 210 W
Line Regulation
Voltage 0.01 %+1 mV
Current 0.01 %+250 uA
Load Regulation
1
Voltage 0.01 %+1 mV
Current 0.01 %+250 uA
Ripple and Noise
(20 Hz – 20 MHz)
Normal Mode
Voltage
0.35 mVrms
/ 3 mVpp
0.5 mVrms /
5 mVpp
0.35 mVrms
/ 3 mVpp
0.5 mVrms /
5 mVpp
0.35 mVrms /
3 mVpp
0.5 mVrms
/5 mVpp
1.5 mVrms /
15 mVpp
4.5 mVrms /
45 mVpp
Normal Mode
Current
2 mA rms
Common Mode
Current
1.5 uA rms
Programming /
Readback Resolution
Voltage 1 mV 2 mV 1 mV 2 mV 1 mV 2 mV 10 mV 20 mV
Current 1 mA 0.1 mA 1 mA 0.1 mA 1 mA 0.2 mA 0.1 mA 0.01 mA
Programming /
Readback Accuracy
±(% output+offset)
Voltage
0.05 %
+5 mV
0.05 %
+10 mV
0.05 %
+5 mV
0.05 %
+10 mV
0.05 %
+5 mV
0.05 %
+10 mV
0.05 %
+50 mV
0.05 %
+100 mV
Current
0.1 %
+2 mA
0.1 %
+1 mA
0.1 %
+2 mA
0.1 %
+1 mA
0.1 %
+2 mA
0.1 %
+5 mA
0.1 %
+2 mA
0.1 %
+1 mA
0.1 %
+0.1 mA
Temperature
Coefficient per °C
±(% output+offset)
Voltage 0.005 %+1 mV
0.005 %
+10 mV
0.005 %
+20 mV
Current 0.01 %+3 mA
Transient Response Time
2
50 µs for output to recover to within 15 mV
100 µs for
output to
recover to within
50 mV
100 µs for
output to
recover to within
120 mV
Settling Time
3
30 ms
Measurement Time 5 ms
OVP Accuracy 0.5 %+0.1 V 0.5 %+1 V
OCP Accuracy 0.5 %+0.1 A
OVP/OCP Response Time
4
1 ms
Rising Time at Full Load / No Load 8 ms 10 ms 8 ms 10 ms 8 ms 10 ms 30 ms 40 ms
Falling Time at Full Load 8 ms 10 ms 8 ms 10 ms 8 ms 10 ms 30 ms 40 ms
Falling Time at No Load 250 ms
Stability (8 hrs)
±(% output+offset)
Voltage 0.02 %+2 mV
0.02 %
+10 mV
0.02 %
+20 mV
Current 0.1 %+1 mA
Command Processing Time
5
10 ms (typical)
Notes:
1 With sense terminal connected.
2 Following a change in output current from full load to half load or vice versa.
3 Maximum time required for the output voltage to change from 1% to 99% or vice versa after receiving a VOLTage or VSET command via GPIB or USB interface.
4 Average time for output to start to drop after OVP/OCP condition occurs.
5 Maximum time for output to change after receipt of commands.