User Manual

70
3. The TRK annunciator will appear on the display to indicate
tracking mode is enabled. Both channels are now
synchronized.
3.7 Remote Sense
Single channel models 9171B, 9172B, 9181B, 9182B, and 9183B
have both front and rear panel remote sense terminals. Dual
channel models 9173B and 9174B, and high voltage single channel
models 9184B and 9185B have rear panel remote sense terminals
only. Remote sense can be used to compensate for voltage drops
due to resistance from test leads connected to your device under
test (DUT), thus providing more accurate output voltage. The power
supply is initially setup to local sense mode by default. Refer to the
following sections for details of local and remote sense setup.
Local Sense
All power supplies are setup with local sense by default. In local
sense mode, shorting bars between the sense terminals and output
terminals are connected on the front panel (if available) and rear
panel outputs. Refer to Figure 16 below (Note: Some models do not
have front panel sense terminals, indicated by +S and -S):
TRK
CH1OFF 10 . 000 V 1 . 000 A
CH2OFF 10 . 000 V 1 . 000 A