User Manual
70 
3.  The TRK annunciator will appear on the display to indicate 
tracking mode is enabled. Both channels are now 
synchronized. 
3.7  Remote Sense 
Single channel models 9171B, 9172B, 9181B, 9182B, and 9183B 
have both front and rear panel remote sense terminals. Dual 
channel models 9173B and 9174B, and high voltage single channel 
models 9184B and 9185B have rear panel remote sense terminals 
only. Remote sense can be used to compensate for voltage drops 
due to resistance from test leads connected to your device under 
test (DUT), thus providing more accurate output voltage. The power 
supply is initially setup to local sense mode by default. Refer to the 
following sections for details of local and remote sense setup. 
Local Sense 
All power supplies are setup with local sense by default. In local 
sense mode, shorting bars between the sense terminals and output 
terminals are connected on the front panel (if available) and rear 
panel outputs. Refer to Figure 16 below (Note: Some models do not 
have front panel sense terminals, indicated by +S and -S): 
TRK  
CH1OFF 10 . 000 V 1 . 000 A 
CH2OFF 10 . 000 V 1 . 000 A 










