User Manual
176 
Figure 32- Current Calibration with Shunt 
Follow the steps below to proceed with current calibration: 
1.  From the calibration menu, press       to enter CURRENT 
CALIBRATION. The following screen will be shown. 
2.  Select the channel (for dual channel models), then press       . 
The calibration process will automatically start. 
CURRENT CALIBRATION CH: 1 
IL Set = X . XXXXA Read = XXX 
IML Set = X . XXXXA Read = XXX 
Enter 
1 
6 ½ Digit DMM 