User Manual
89 
1.  Press          , then    to enter SYSTEM SETTING.  
2.  Go down to EXTERN CONTROL (this option is only 
available with DIO/Analog Card installed) and press   or 
 to select between OFF, VOLT, or RES. Select VOLT 
for external voltage control or RES for external resistance 
control. Select OFF to disable external analog control. 
3.  Press       to save the change. If VOLT is selected in the 
previous step, go to EXTERN LEVEL setting right below 
EXTERN CONTROL and press   or   to select 
between 5V or 10V. Selecting 5V means a 0 – 5V source 
will be used to control the full range of the voltage/current. 
Similarly, 10V means a 0 – 10V source will be used. 
4.  Press        again to save the change, and press           twice 
to go to the main screen, which will look like the following 
below (Note that voltage and current resolution will be 10 mV 
and 10 mA for external analog control): 
For Dual Channel models: 
OFF 
0.00 V 0.00 A 
Menu 
1 
Enter 
Enter 
Esc 