User Manual
181 
Figure 35- Dual Channel External Analog Input Calibration 
2.  Connect the 10VDC source to the analog inputs on the 
DIO/Analog Interface Card as shown in Figure 36 for single 
channel. 
Figure 36- Single Channel Analog Control Calibration Setup 
3.  Set the DC source to output 1.000 V and wait for approximately 
5 seconds until the output voltage is stable. Press       to finish 
the first step. 
4.  Next, change the output of the DC source to 9.000 V and follow 
the same as step 3. 
Calib EXT . Lo Volt = 1 . 000 V 
Calib EXT . Lo Volt = 9 . 000 V 
CH 1 EXT . VOLT Res = Start 
CH 1 EXT . CURR Res = Start 
CH 2 EXT . VOLT Res = Start 
CH 2 EXT . CURR Res = Start 
Enter 