User Manual
177 
3.  The power supply will output a pre-set current to the DMM. Wait 
for the DMM reading to settle, then enter it into where it shows 
IL Set = __.____A on the display. Press       to proceed to IML. 
If current shunt is used for the calibration, divide the DMM 
readings (in DCV) by the resistance (Rs) of the shunt. The result 
should be DCV/Rs.  Enter this current result into IL. 
4.  The power supply will then pre-set another current value for IML. 
Repeat the same process in step 3. Once finished, repeat the 
same process for IMH and IH, as shown below. 
5.  Once finished, the OCP Lo calibration will begin. Press              
to start. After it’s complete, the unit will select CALIB OCP Hi for 
calibration. Again, after it’s complete, press        . 
6.  Once this calibration process is complete, the calibration data 
will be stored into flash memory, and the supply will return to the 
calibration menu. 
IMH Set = X . XXXXA Read = XXXXX 
IH Set = X . XXXXA Read = XXXXX 
CALIB OCP Lo = Start 
CALIB OCP Hi =  Start 
Enter 
Enter 
Enter 