The following equipment is required for the calibration procedure: 
  6 ½ digit precision digital multimeter 
  Shunt  for  current  calibration  (Necessary  if  the  maximum 
output  current  of  the  power  supply  is  greater  than  the 
maximum current the multimeter is capable of measuring) 
  DC source with an output up to 10VDC and accuracy of at 
least ±0.1% (for external analog input calibration) 
  5 kΩ resistor with accuracy within ±0.1% or better (for external 
analog input calibration) 
Current Calibration 
-  If the maximum output current of the power supply is within the 
current limit of the DMM, then use the wire connections shown in 
Figure 31 to calibrate current. 
-  If the maximum output current of the power supply is greater than 
the current limit of the DMM, then a current shunt must be used. 
The wire connections between the power supply and the DMM will 
be setup like Figure 32. 
Figure 31 - Current Calibration Setup 
6 ½ Digit DMM 