User Manual

54
5. Select Usminand press enter. The user can choose a Usmin value between 1 V 80 V.
Once a value is selected, press to confirm.
6. Use the key to select Off/On and use keys to toggle between the two options.
Choose On to be able to run the function.
7. Once both parameters are set, press to confirm the changed settings and press
several times to exit the menu. If the Starting profile function was setup properly,
the Trig indicator will be highlighted and Profile for the reset test will be shown on the
front panel, shown below.
8. To enable the Profile reset test function, press the key, on the front panel. The
proper voltage selected should be metered on the front panel display.
9. To run the Profile reset test function press and to trigger the test to run.
Please be aware this test does run for 340 seconds (See profile reset test figure above).
Starting profile
There are different profiles available for 12 V systems and 24 V systems, please refer to the
tables above to configure the power supply options properly.
Follow the steps below to set up the ISO16750-2 for Starting profile function:
1. Access the main menu by pressing and the .
2. Use the key to select FUNC. and press .
3. Use the key to select ISO16750-2 and press . The screen below will display:
4. While Short voltage drop is flashing, use keys to toggle to Starting profile (shown
below) and press to access that option.
0.000V 0.000A
0.0W Profile for reset
Trig
ISO16750-2 profile
Profile for the reset test