User Manual
54 
5.  Select Usmin… and press enter. The user can choose a Usmin value between 1 V – 80 V. 
Once a value is selected, press  to confirm. 
6.  Use the   key to select Off/On and use     keys to toggle between the two options. 
Choose On to be able to run the function. 
7.  Once both parameters are set, press  to confirm the changed settings and press 
several times to exit the menu. If the Starting profile function was setup properly, 
the Trig indicator will be highlighted and Profile for the reset test will be shown on the 
front panel, shown below. 
8.  To enable the Profile reset test function, press the   key, on the front panel. The 
proper voltage selected should be metered on the front panel display. 
9.  To run the Profile reset test function press   and  to trigger the test to run. 
Please be aware this test does run for 340 seconds (See profile reset test figure above).  
Starting profile 
There are different profiles available for 12 V systems and 24 V systems, please refer to the 
tables above to configure the power supply options properly. 
Follow the steps below to set up the ISO16750-2 for Starting profile function: 
1.  Access the main menu by pressing   and the . 
2.  Use the    key to select FUNC. and press .  
3.  Use the    key to select ISO16750-2 and press . The screen below will display: 
4.  While Short voltage drop is flashing, use  keys to toggle to Starting profile (shown 
below) and press   to access that option.  
0.000V  0.000A 
 0.0W      Profile for reset   
Trig 
ISO16750-2 profile 
Profile for the reset test 