User Manual
49
Table 3.4 – Parameters for test pulse
Follow the steps below to set up the DIN40839 function:
1. Access the main menu by pressing and the .
2. Use the key to select FUNC. and press .
3. Use the key to select DIN40839 and press . The screen below will display:
4. While 12V or 24V is flashing, use keys to toggle between the two options. Set to
12V for the 12V test function and 24V for the 24V test function.
5. Use the key to select Off/On and use keys to toggle between the two options.
Choose On to be able to run the function.
6. Once both parameters are set, press to confirm the changed settings and press
twice to exit the menu. If the DIN40839 function was setup properly, the Trig
indicator will be highlighted and DIN40839 will be shown on the front panel, shown
below.
DIN40839
12V Off