User Manual
47 
1.  From the LIST menu, use the key to select Recall and press  . The display will 
say RECALL LIST with the prompt Recall List File: to ask to enter a number. 
2.  This is the location number where the program is stored in. Select a number between 1 
to 10 with the current adjust knob or the numeric keypad, then press  . 
3.  The display will return back to the LIST menu. 
Run Program 
To run a program sequence after it is recalled, the trigger source must be configured first. Refer 
to “Configure Trigger Source” in section 3.3 to set up the trigger source. 
Follow the steps below to run the program: 
1.  Assuming the stored program has already been recalled, from the LIST menu, with Off 
selected, press  . The Trig indicator will appear at the top middle of the display. 
2.  Now, press   twice to exit the menu. The following will display: 
3.  If the display is showing the measured voltage and current value, the   backlight 
should be lit and display should show 0.000V and 0.000A. 
4.  Now, press the   key to enable (ON) the output. 
5.  If the Trigger Source configured in section 3.3 is set to Manual (Def), then the front 
panel trigger key can be used to initiate the program. Press   and then . 
6.  The program will immediately start running and the backlight of the   button will 
be blinking. The Trig indicator will also disappear. 
7.  When the program ends, the   will stop blinking and the Trig indicator will 
reappear again. 
Note: If the Trigger Source is configured to BUS, use the remote command for sending 
triggers over any of the remote interface available. Refer to the programming 
manual for details. 
0.000V 
 0.000A 
0.0W 
 OFF        Trig   
LIST 