9.  This is prompting the user to enter the voltage setting for step 1. Use the voltage and 
current adjust knob or the numeric keypad and enter 1.000. Then press . 
10. The display will change to the following, prompting the user to enter the current setting 
for step 1. 
11. Similar to entering the voltage setting in the previous step, do the same to enter 1.000 
as the current setting. Then press . Now, the following will display: 
12. Enter the width, which is the time to hold the voltage and current setting configured in 
the previous steps.  Enter 1.000 (for 1 second). Press . Now, the following will 
display: 
13. Enter the slope, which defines the voltage slew rate (transition time between steps) for 
step 1. Enter 0.005. Press . 
14. Now, the display will return back to the display shown in step number 8, except it will 
prompt the user to enter the voltage setting for step 2 of the sequence. Follow steps 8-
13 to configure step 2 with: 2V, 1A, 2s width, 0.005s slope. Then step 3 with: 3V, 1A, 3s 
width, 0.005s slope. Then step 4 with: 1.5V, 1A, 3s width, 0.005s slope. Finally, step 5 
with: 2.5V, 1A, 4s width, 0.005s. 
15. After step 5 parameters are configured, the display will show the following to request 
confirmation to save the step parameters into memory. Press  key to select Yes and 
press . 
EDIT SEQ SLOPE 
Seq Step 1 Slope =  0.005s 
EDIT SEQ WIDTH 
Seq Step 1 Width = 1.000s 
EDIT SEQ CURRENT 
Seq Step 1 Current = 1.000A 