User Manual
28 
Note:  The default is 4800, 8, N, 1, Address = 0.  
*Setting the baudrate to 115200 may provide unstable results during remote 
communication. Select a lower baudrate if communication errors occur. 
6.  Select Addr… and press . You will be prompted to enter an address. Use current 
adjust knob or numeric keypad to enter a number between 0 to 31, and then press
. 
7.  For each power supply that you want to control, provide a different address. For 
example, if you have three power supplies to control, set the first supply to address 1, 
second supply to address 2, and third supply to address 3. If two or more supplies have 
the same address, the RS485 protocols will create communication errors. 
8.  Repeat the above steps for each power supply that you want to connect together and 
control, making sure that each of them have a different address assigned. 
9.  Press   several times to exit the menu. 
Ethernet (Model 9117 only) 
The Ethernet (LAN) interface supports a raw socket TCP/IP connection only. It does not support 
UDP protocols. 
Follow the below instructions to select Ethernet interface for remote operation. 
1.  From the SYSTEM menu, browse and select Communication and press . 
2.  Select Ethernet and press   to set to Ethernet for remote communication. 
3.  Use  to cycle through the Ethernet settings. 
4.  Use the current adjust knob or the numeric keypad to change the desired address or 
value for each Ethernet settings. 
5.  Below lists the options that can be changed for each setting: 
Gateway: 192.168.0.1 
IP: 192.168.0.125 
Mask: 255.255.255.0 
Socket Port: 30000 
6.  Press   to save the selected address and the display will return to the 
Communication menu. 
7.  Press   several times to exit the menu. 