User Manual

Specifications
Note: All specifications apply to the unit after:
Temperature Stabilization time: 30 mins
Operating Temperature: 23 °C ± 5 °C
Specifications are subject to change without notice.
Test Signal Frequency
Model
Range
Minimum
resolution
Accuracy
894
20 Hz - 500 kHz
0.01 Hz
0.01 %
895
20 Hz - 1 MHz
Test Signal Levels
AC source (ALC* OFF)
Voltage Accuracy
10% x set voltage ± 2mV
Voltage Level
Resolution
5 mVrms - 100 mVrms
100 µVrms
100 mVrms - 1 Vrms
1 mVrms
1 Vrms - 2 Vrms
10 mVrms
Current Accuracy
10 % x set current ± 10 µA
Current Range
Impedance
166.7 µArms - 66.7 mArms
30
100.0 µArms - 40.0 mArms
50
50.0 µArms - 20.0 mArms
100
AC source (ALC* ON)
1
Voltage
Range
10 mVrms 1 Vrms
Accuracy
6 % x set voltage ± 2 mV
Current
Range
100 µArms - 10 mArms
Accuracy
6 % x set current ± 10 µA
DC bias source
Voltage
Range
-5 V to +5 V
Accuracy
1 % x set voltage ± 5 mV
Resolution
0.01 mV
Current
Range
-50 mA to +50 mA
Accuracy
1 % x set current ± 50 µA