Limit Table Setup
The Limit Table Setup allows the user to set the parameters for the Bin Sorting Function. This
page allows the user to set the parameters for the built-in comparator. These LCR meters can
set nine bin limits of the primary parameters and one of the secondary parameters. The tested
result can be divided into up to ten bins (BIN 1 to BIN 9 and BIN OUT). If the primary parameter
of the DUT (device under test) is within the limit range from BIN1 to BIN9, but the secondary
parameter is out of the limit range, the DUT can be sorted into aux bin.
Press SETUP and then LIMIT TABLE to enter into the <LIMIT TABLE SETUP> page as the following
figure shown.
Figure 19 - Limit Table Setup
The following limit parameters of the Compare Function can only be set on <LIMIT TABLE
SETUP> page.
Measurement parameters (PARAM)
Compare function modes (MODE)
Low and High values for the Secondary Parameter
Low and High values for the Secondary Parameter