User Manual
46
Frequency
(kHz)
Capacitors
Inductors
Use series
model
when
Use
parallel
model
when
Use series
model
when
Use
parallel
model
when
0.1
> 16 μF
< 16 μF
< 160 mH
> 160 mH
1
> 1.6 μF
< 1.6 μF
< 16 mH
> 16 mH
10
> 160 nF
< 160 nF
< 1.6 mH
> 1.6 mH
100
> 16 nF
< 16 nF
> 160 μH
> 160 μH
Table 10 - Series/Parallel Model Guideline
Measure Setup Menu
The Measure Setup page allows the user modify a series of parameters used while
performing a measurement. To access the Measure Setup menu, press the SETUP up
button. The screen display will look like Figure 18.
Figure 18 - Measure Setup Menu
In this page, the following control parameters can be set: Trigger Mode (TRIG)
Auto Level Control (ALC)
Delay Time (DELAY)
Output Impedance (Rsou.)
Average times (AVG)
Voltage Level Monitor ON/ OFF (Vm)