User Manual
37 
DCR OPEN: Tests the Open resistance for DCR. 
The following soft key will be displayed on the top right corner of the display: 
  ABORT: This soft key cancels the current open correction operation. 
4)  Keep  the  test  fixture  open  (no  connection),  then  press  the  softkey  MEAS  OPEN  to 
execute open calibration. The meter will beep when the correction is completed.  
Short Correction 
The short correction feature compensates for any residual impedance that may exist within the 
test fixture or leads, as shown in Figure 12. Short Correction uses 48 fixed frequencies to adjust 
and compensate the input.  
Figure 12 – Residual Inductance 
Short Correction Procedure 
1)  Connect the test fixture to the test terminal, or insert a shorting bar into the test 
fixture. 
2)  Press the SETUP button and enter the MEASURE SETUP menu.  
3)  Press  the  softkey,  CORRECTION,  to  enter  the  Correction  menu.  Press  the  Down 
Key and select the SHORT option. The following soft keys will be displayed on the 
soft keys to the right of the screen: 
ON:  Applies a SHORT correction adjustment to the current measurement.  
Press ON to perform a short-circuit correction.  
OFF:  Turns off the SHORT correction. 
Press  OFF  to  turn  the  SHORT  correction  function  off.  There  will  be  no  Short 
Correction performed in measurements. 
MEAS SHORT: Tests the residual impedance (resistance and reactance) over the 48 pre-
set frequencies when shorted. 