User Manual
32 
  ↓(-) - This soft key is used to decrease the output level of DC bias. 
1)  Use soft keys or numeric keys to set the DC bias source. When using numeric keys, the 
soft key displays the available units (mV, V, µA, mA and A). You can use them to input 
the unit after entering the numeric value. If the [ENTER] button is pressed to input bias 
value, the unit defaults to V or A. 
Press the [DC BIAS] key on the front panel to enable DC bias output. The [DC BIAS] key will 
be lid. 
Test speed 
The test speed is determined by the following factors: 
  Integration time (A/D conversion) 
  Average test times (average test times per test) 
  Measurement delay (from startup to the start of measurement) 
  Display time of test results 
You can select FAST, MED or SLOW. Generally, test results are more stable and accurate when 
set to SLOW.  
  Fast: Approx. 75 times/s (13 ms/reading) 
  Medium: Approx. 11 times/s (90 ms/time) 
  Slow: Approx. 2.7 times/s (370 ms/time) 
Note: The fast and middle speed will be slow down when frequency <10 kHz. 
Setting the test speed: 
Move the cursor to SPEED, and the following soft keys will be displayed: 
  FAST 
  MED 
  SLOW 
Use the soft keys to select the test speed. 
Digits Resolution 
The measurement results of the meter can display up to 6 floating-point digits. The 894 and 
895 can be manually configured to display a fixed number of digits or automatically select 
the number of digits based on range. 