List Sweep Comparison Signals  
The List Sweep Comparison signal output is different from other modes, including Bin 
Comparison: 
  List Sweep Comparison Output Signal 
/BIN1-/BIN9 and /OUT show the IN (good) or /OUT (over the standard) result.   
/AUX shows PASS/FAIL result. 
When a sweep test is finished, these signals will be at the output signal. 
  Control Output Signal 
/INDEX (Analog test finished) 
 /EOM (End of measurement). 
  SEQ sweep mode 
/INDEX is defined as the effective signal when the last sweep point of the analog test is 
finished.  
/EOM is defined as the effective signal when all the test results are effective after every list 
sweep task is finished. 
  STEP sweep mode 
/INDEX is defined as the effective signal when the analog test of every sweep point is 
finished.  
/EOM is defined as the effective signal when the test and the comparison of every step are 
finished.  
The pin distribution for list sweep is shown in the table below and in Figure 33. The timing 
chart is shown in Figure 36.  
The table below shows the list sweep comparison pin assignment and functionality: 
Pin 
Signal 
Description 
1 
2 
3 
4 
5 
6 
7 
8 
9 
10 
/BIN1 
/BIN2 
/BIN3 
/BIN4 
/BIN5 
/BIN6 
/BIN7 
/BIN8 
/BIN9 
/OUT 
Out of the limit of sweep point1 
Out of the limit of sweep point2 
Out of the limit of sweep point3 
Out of the limit of sweep point4 
Out of the limit of sweep point5 
Out of the limit of sweep point6 
Out of the limit of sweep point7 
Out of the limit of sweep point8 
Out of the limit of sweep point9 
Out of the limit of sweep point10 
11 
/AUX 
/AUX is asserted when at least one device under test is not good in the 
list. 